| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
056
|
Date of Entry 06/07/2021
|
|
FR Recognition Number
|
11-382
|
| Standard | |
ASTM F3090-20 Standard Test Method for Fatigue Testing of Acetabular Devices for Total Hip Replacement |
|
Scope/Abstract1.1 This test method is intended to evaluate the fatigue strength of metallic acetabular shells with hemispheric outer surfaces.
1.2 This test method, as described, is not intended to evaluate the following: the strength of components that may be mated with the acetabular shells (for example, augments, acetabular liners); attributes of the shells not related to strength (for example, fixation, coating adhesion); the strength of acetabular shell features away from, or loaded differently than, the primary load bearing region of the shell (for example, screws, spikes, flanges); non-hemispherical shells (for example, patient-matched geometries); or corrosion between modular components.
1.3 Modifications to this test method (for example, different support medium, different /position of unsupported region, different testing environment) may result in a method appropriate to evaluate the characteristics listed in 1.2. Such modification must have adequate justification.
1.4 Although the methodology described does not replicate all physiological force conditions, it is a means of in vitro comparison of acetabular device designs under repeated forces. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM F3090-20 [Rec# 11-382] will be superseded by recognition of ASTM F3090-24 [Rec# 11-417]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-382] until December 20, 2026. After this transition period, declarations of conformity to [Rec# 11-382] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §888.3300 |
Prosthesis, Hip, Constrained, Metal
|
Class 3
|
KXD
|
| §888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer
|
Class 2
|
KWZ
|
| §888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer, + Additive
|
Class 2
|
PBI
|
| §888.3320 |
Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)
|
Class 3
|
JDL
|
| §888.3330 |
Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
|
Class 3
|
KWA
|
| §888.3340 |
Prosthesis, Hip, Semi-Constrained, Composite/Metal
|
Class 2
|
KMC
|
| §888.3350 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
|
Class 2
|
JDI
|
| §888.3350 |
Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented
|
Class 2
|
OQH
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
|
Class 2
|
LZO
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish
|
Class 2
|
MAY
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
|
Class 2
|
MEH
|
| §888.3353 |
Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented
|
Class 2
|
OQI
|
| §888.3358 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
|
Class 2
|
LPH
|
| §888.3358 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
|
Class 2
|
MBL
|
| §888.3358 |
Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented
|
Class 2
|
OQG
|
| §888.3410 |
Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer
|
Class 3
|
KXB
|
| §888.3410 |
Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer, Uncemented
|
Class 3
|
OCG
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented |
Class 3
|
LPF
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented |
Class 3
|
MRA
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing |
Class 3
|
NXT
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Ceramic-On-Metal Articulation |
Class 3
|
OVO
|
|
Relevant FDA Guidance and/or Supportive Publications*
| Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |
|
|