• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 049 Date of Entry 06/07/2018 
FR Recognition Number 6-406
Standard
ASTM  F1862/F1862M-17
Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Known Velocity)
Scope/Abstract
1.1 This test method is used to evaluate the resistance of medical face masks to penetration by the impact of a small volume (~2 mL) of a high-velocity stream of synthetic blood. Medical face mask pass/fail determinations are based on visual detection of synthetic blood penetration.
1.2 This test method does not apply to all forms or conditions of blood-borne pathogen exposure. Users of the test method must review modes for face exposure and assess the appropriateness of this test method for their specific application.
1.3 This test method primarily addresses the performance of materials or certain material constructions used in medical face masks. This test method does not address the performance of the medical face masks design, construction, or interfaces or other factors with the potential to affect the overall protection offered by the medical face mask and its operation (such as filtration efficiency and pressure drop). Procedures for measuring these properties are contained in Test Method F2101 and MIL-M-36954C.
1.4 This test method does not address breathability of the medical face mask materials or any other properties affecting the ease of breathing through the medical face mask. This test method evaluates medical face masks as an item of protective clothing. This test method does not evaluate the performance of medical face masks for airborne exposure pathways or in the prevention of the penetration of aerosolized body fluids deposited on the medical face mask.
1.5 The values stated in SI units or inch-pound units are to be regarded separately as standard. The pressure values stated in each system are not exact equivalents. However, as the corresponding velocities are within 1 % of each other, (see X1.4.2), reporting of the results in either units is permitted.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.

Note: Subclause 9.2.1 states a single sampling plan providing an AQL of 4.0% would require 32 specimens (i.e., complete face masks). If less than 32 specimens are tested, appropriate justification would be needed in the report.
Transition Period
FDA recognition of ASTM F1862/F1862M-17 [Rec# 6-406] will be superseded by recognition of ASTM F1862/F1862M-24 [Rec# 6-504]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 6-406] until December 20, 2026. After this transition period, declarations of conformity to [Rec# 6-406] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§878.4040 Mask, Surgical Class 2 FXX
§878.4040 Respirator, Surgical Class 2 MSH
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submissions; Guidance for Industry and FDA, issued March 2004.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Yongqing Chen
  FDA/OC/CDRH/OPEQ/OHTIV/DHTIVC/
  240-402-9433
  Yongqing.Chen@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
General Plastic Surgery/General Hospital
*These are provided as examples and others may be applicable.
-
-