| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
064
|
Date of Entry 05/26/2025
|
|
FR Recognition Number
|
17-21
|
| Standard | |
ASTM F647-22 Standard Practice for Evaluating and Specifying Implantable Shunt Assemblies for Neurosurgical Application |
|
Scope/Abstract1.1 This practice covers requirements for the evaluation and specification of implantable shunts as related to resistance to flow, direction of flow, materials, radiopacity, mechanical properties, finish, sterility, and labeling of shunt assemblies.
1.2 Devices to which this practice is applicable include, but are not limited to, those that are temporarily implanted to effect external drainage; or permanently implanted to effect shunting of fluid from a cerebral ventricle, a cyst, the subarachnoid space to the peritoneal cavity, the venous circulation, or some other suitable internal delivery site, and intracranial bypass.
1.3 Limitations - Although this practice includes a standard test method for the evaluation of pressure/flow characteristics of shunts or shunt components, it does not include specific pressure/flow requirements.
1.4 The following components that individually or in combination comprise shunt assemblies are considered to be within the scope of this practice: catheters (such as atrial, peritoneal, ventricular), connectors, implantable accessory devices (such as antisiphon devices and reservoirs), valved catheters, and valves.
NOTE 1 - The standards in Section 2 contain provisions that, through reference in this text, constitute provisions of this practice. At the time of publication, the editions indicated are valid. All standards are subject to revision, and parties to agreements based on this practice are encouraged to investigate the possibility of applying the most recent editions of the standards indicated below. Devices or components, or both, whose structures are comparable to that outlined in these standards are acceptable. |
|
Extent of Recognition
| Partial recognition. The following part(s) of the standard is (are) not recognized: |
* Subclause 5.3 Biocompatibility (Polymeric Materials)
* Subclause 6.1.4 Biological Properties
* Appendix X2.3 and X2.4 of X2 Guidance on Materials
|
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
This standard is recognized in part because:
* Subclause 5.3 is in conflict with existing published final guidance, see Section III, IV, and Table A.1 of Ref #1 listed below.
* Subclause 6.1.4 is in conflict with existing published final guidance, see Section III, IV, and Table A.1 of Ref #1 listed below.
* Appendix X2.3 and X2.4 are in conflict with existing published final guidance, see Section III, IV, and Table A.1 of Ref #1 listed below. |
|
Transition Period
| FDA recognition of ASTM F647-94 (Reapproved 2014) [Rec# 17-4] will be superseded by recognition of ASTM F647-22 [Rec# 17-21]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 17-4] until July 4, 2027. After this transition period, declarations of conformity to [Rec# 17-4] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §882.4060 |
Cannula, Ventricular
|
Class 1
|
HCD
|
| §882.4100 |
Catheter, Ventricular
|
Class 2
|
HCA
|
| §882.4545 |
Instrument, Shunt System Implantation
|
Class 1
|
GYK
|
| §882.5550 |
Shunt, Central Nervous System And Components
|
Class 2
|
JXG
|
|
Relevant FDA Guidance and/or Supportive Publications*
1. Use of International Standard ISO 10993-1, "Biological Evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," Guidance for Industry and Food and Drug Administration Staff, Issued September 2023.
2. Guidance for Industry: Pyrogens and Endotoxins Testing: Questions and Answers, Issued June 2012.
3. AAMI/ANSI ST72:2011/(R)2016, Bacterial Endotoxins - Test Methods, Routine Monitoring, and Alternatives to Batch Testing.
4. AAMI/ANSI/ISO 10993-1:2009/(R)2013, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process.
5. USP-NF <85> Bacterial endotoxins test.
6. USP-NF <151> Pyrogen test.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|