Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 064 Date of Entry 05/26/2025 
FR Recognition Number 9-151
Standard
ASTM F1828-22
Standard Specification for Ureteral Stents
Scope/Abstract
1.1 This specification covers the referee test methods for evaluating the performance characteristics of a single-use ureteral stent with retaining means at both ends, during short-term use for drainage of urine from the kidney to the bladder. These stents are typically available in diameters of 3.7 Fr to 14.0 Fr, and lengths of 8 cm to 30 cm, and are made of silicone, polyurethane, and other polymers. They are provided non-sterile for sterilization and sterile for single use.

1.2 Exclusions-Long-term indwelling usage (over 30 days) is encountered with this product, but not commonly, and is therefore considered an exception to this specification. Similarly, the use of ureteral stents for non-ureteral applications such as nephrostomy and ileostomy is excluded from the scope of this specification. Non-sterile ureteral stents are also excluded due to the variability of hospital sterilization equipment and processes and the resulting effects on ureteral stent characteristics.
...
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§876.4620 Stent, Ureteral Class 2 FAD
Relevant FDA Guidance and/or Supportive Publications*
Guidance for the Content of Premarket Notifications for Ureteral Stents, Issued February, 1993.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Angel Soler-Garcia
  FDA/OC/CDRH/OPEQ/OHTIII/DHTIIIB/
  301-796-6535
  Angel.Soler-Garcia@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
ObGyn/Gastroenterology/Urology
*These are provided as examples and others may be applicable.
-
-