| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
064
|
Date of Entry 05/26/2025
|
|
FR Recognition Number
|
11-420
|
| Standard | |
ASTM F2580-24 Standard Test Method for Evaluation of Modular Connection of Proximally Fixed Femoral Hip Prostheses |
|
Scope/Abstract1.1 This test method covers a procedure for the fatigue testing of metallic femoral hip prostheses used in hip joint replacements. This test method covers the procedures for the performance of fatigue tests on metallic femoral hip stems using a cyclic, constant-amplitude force. It applies to hip prostheses that utilize proximal metaphyseal fixation and are of a modular construct, and it is intended to evaluate the fatigue performance of the modular connections in the metaphyseal filling (that is, proximal body) region of the stem.
1.2 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic hip prostheses while held in a proximally fixated manner, with the distal end not held by a potting medium. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM F2580-18 [Rec# 11-344] will be superseded by recognition of ASTM F2580-24 [Rec# 11-420]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-344] until July 4, 2027. After this transition period, declarations of conformity to [Rec# 11-344] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §888.3300 |
Prosthesis, Hip, Constrained, Metal
|
Class 3
|
KXD
|
| §888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer
|
Class 2
|
KWZ
|
| §888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer, + Additive
|
Class 2
|
PBI
|
| §888.3320 |
Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)
|
Class 3
|
JDL
|
| §888.3330 |
Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
|
Class 3
|
KWA
|
| §888.3340 |
Prosthesis, Hip, Semi-Constrained, Composite/Metal
|
Class 2
|
KMC
|
| §888.3350 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
|
Class 2
|
JDI
|
| §888.3350 |
Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented
|
Class 2
|
OQH
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
|
Class 2
|
LZO
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish
|
Class 2
|
MAY
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
|
Class 2
|
MEH
|
| §888.3353 |
Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented
|
Class 2
|
OQI
|
| §888.3358 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
|
Class 2
|
LPH
|
| §888.3358 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
|
Class 2
|
MBL
|
| §888.3358 |
Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented
|
Class 2
|
OQG
|
| §888.3360 |
Prosthesis, Upper Femoral
|
Class 2
|
JDD
|
| §888.3360 |
Prosthesis, Hip, Femoral Component, Cemented, Metal
|
Class 2
|
JDG
|
| §888.3360 |
Prosthesis, Hip, Hemi-, Femoral, Metal
|
Class 2
|
KWL
|
| §888.3360 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented
|
Class 2
|
LWJ
|
| §888.3390 |
Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
|
Class 2
|
KWY
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented |
Class 3
|
LPF
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented |
Class 3
|
MRA
|
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff - Non-clinical Information for Femoral Stem Prostheses, issued September 2007.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |
|
|