| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
064
|
Date of Entry 05/26/2025
|
|
FR Recognition Number
|
11-422
|
| Standard | |
ASTM F1798-24 Standard Test Method for Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants |
|
Scope/Abstract1.1 This test method covers the measurement of uniaxial static and fatigue strength, and resistance to loosening of the component interconnection mechanisms of spinal arthrodesis implants.
1.2 The purpose of this test method is to provide a means of mechanically characterizing different designs of spinal implant interconnections. Ultimately, the various components and interconnections should be combined for static and fatigue testing of the spinal implant construct. It is not the intention of this test method to address the analysis of spinal implant constructs or subconstructs or to define levels of performance of spinal implants, as insufficient knowledge is available to predict the consequences of the use of particular spinal implant designs.
1.3 This standard defines test methods to measure the strength of spinal implant component interconnections and how to report test results. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM F1798-21 [Rec# 11-402] will be superseded by recognition of ASTM F1798-24 [Rec# 11-422]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-402] until July 4, 2027. After this transition period, declarations of conformity to [Rec# 11-402] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §888.3050 |
Appliance, Fixation, Spinal Interlaminal
|
Class 2
|
KWP
|
| §888.3060 |
Appliance, Fixation, Spinal Intervertebral Body
|
Class 2
|
KWQ
|
| §888.3070 |
Thoracolumbosacral Pedicle Screw System
|
Class 2
|
NKB
|
| §888.3070 |
Posterior Metal/Polymer Spinal System, Fusion
|
Class 2
|
NQP
|
| §888.3070 |
Growing Rod System
|
Class 2
|
PGM
|
| §888.3070 |
Growing Rod System- Magnetic Actuation
|
Class 2
|
PGN
|
| §888.3075 |
Posterior Cervical Screw System
|
Class 2
|
NKG
|
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance for Industry and FDA Staff - Spinal System 510(k)s, issued May 2004.
Guidance for Industry and/or FDA Reviewers/Staff - Guidance Document for the Preparation of IDEs for Spinal Systems, issued January 2000.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |