| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
052
|
Date of Entry 07/15/2019
|
|
FR Recognition Number
|
14-528
|
| Standard | |
ISO 11137-1 First edition 2006-04-15 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)] |
|
Identical AdoptionANSI AAMI ISO 11137-1:2006/(R)2015 Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2019)] |
|
Scope/AbstractISO 11137-1:2006 specifies requirements for the development, validation and routine control of a radiation sterilization process for medical devices. Although the scope of ISO 11137-1:2006 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other products and equipment. ISO 11137-1:2006 covers radiation processes employing irradiators using the radionuclide 60Co or 137Cs, a beam from an electron generator or a beam from an X-ray generator. ISO 11137-1:2006 does not: -specify requirements for development, validation and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease; -detail specified requirements for designating a medical device as sterile; -specify a quality management system for the control of all stages of production of medical devices; -specify requirements for occupational safety associated with the design and operation of irradiation facilities; -specify requirements for the sterilization of used or reprocessed devices. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ISO 11137-1 First edition 2006-04-15 [Rec# 14-528] will be superseded by recognition of ISO 11137-1 Second edition 2025-04 [Rec# 14-611]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 14-528] until July 4, 2027. After this transition period, declarations of conformity to [Rec# 14-528] will not be accepted. |
|
Relevant FDA Guidance and/or Supportive Publications*
1. AAMI TIR17:2017 Compatibility of materials subjected to sterilization.
2. AAMI TIR35:2016 Sterilization of health care products - Radiation sterilization - Product adoption and alternative sampling plans for verification dose experiments and sterilization dose audits.
3. AAMI TIR37:2013 Sterilization of health care products - Radiation - Guidance on sterilization of human tissue-based products.
4. AAMI TIR40:2018 Sterilization of health care products - Radiation - Guidance on dose setting utilizing a Modified Method 2.
5. AAMI TIR76:2021 Sterilization of health care products - Radiation - Substantiation of a selected sterilization dose at a specified sterility assurance level: Method Vdmax SD-S.
6. ASTM E3270-21 Standard Guide for Operational Qualification of Gamma Irradiators.
7. ISO/ASTM 51631-20 Practice for use of calorimetric dosimetry systems for dose measurements and dosimetry system calibration in electron beams
8. ISO/ASTM 53239-21 Standard Guide for Using Statistical Process Control Principles for Routine Dosimetry in Radiation Processing.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contacts
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |