| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
065
|
Date of Entry 12/22/2025
|
|
FR Recognition Number
|
5-146
|
| Standard | |
ISO 80369-6 Second edition 2025-05 Small bore connectors for liquids and gases in healthcare applications - Part 6: Connectors for neuraxial applications |
|
Scope/AbstractThis document specifies requirements for small-bore connectors intended to be used for connections in neural applications.
This document does not specify requirements for the medical devices or accessories that use these connectors. Such requirements are given in particular standards for specific medical devices or accessories. |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ISO 80369-6 First Edition 2016-03-15 [Rec# 5-108] will be superseded by recognition of ISO 80369-6 Second edition 2025-05 [Rec# 5-146]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 5-108] until December 17, 2028. After this transition period, declarations of conformity to [Rec# 5-108] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §868.5120 |
Catheter, Conduction, Anesthetic
|
Class 2
|
BSO
|
| §868.5140 |
Anesthesia Conduction Kit
|
Class 2
|
CAZ
|
| §868.5150 |
Needle, Conduction, Anesthetic (W/Wo Introducer)
|
Class 2
|
BSP
|
| §880.5440 |
Set, Administration, Intravascular
|
Class 2
|
FPA
|
| §880.5440 |
Administrations Sets With Neuraxial Connectors
|
Class 2
|
PWH
|
| §880.5860 |
Syringe, Piston
|
Class 2
|
FMF
|
| §882.1620 |
Device, Monitoring, Intracranial Pressure
|
Class 2
|
GWM
|
| §882.4060 |
Cannula, Ventricular
|
Class 1
|
HCD
|
| §882.4100 |
Catheter, Ventricular
|
Class 2
|
HCA
|
| §882.4100 |
Catheter, Ventricular (Containing Antibiotic Or Antimicrobial Agents)
|
Class 2
|
NHC
|
| §882.5550 |
Shunt, Central Nervous System And Components
|
Class 2
|
JXG
|
| §882.5560 |
External Cerebrospinal Fluid (Csf) Diversion
|
Class 2
|
PCB
|
|
Relevant FDA Guidance and/or Supportive Publications*
(Website) Medical Device Connectors: https://www.fda.gov/medical-devices/general-hospital-devices-and-supplies/medical-device-connectors
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
|
|
| *These are provided as examples and others may be applicable. |