| Part B: Supplementary Information Sheet (SIS) |
|
FR Recognition List Number
|
065
|
Date of Entry 12/22/2025
|
|
FR Recognition Number
|
11-437
|
| Standard | |
ASTM F2033-25 Standard Specification for Hip Joint Replacement Bearing Surfaces |
|
Scope/Abstract1.1 This specification covers the requirements for the mating bearing surfaces of total hip joint replacements and separate requirements for dual mobility hip joint replacements, intended for total hip arthroplasty; and partial hip joint replacements and separate requirements for bipolar femoral hip joint replacements, intended for hemiarthroplasty. All joint replacement implant types are defined in ISO 21535. More specifically, this specification covers hip joint replacement of the ball-and-socket configuration.
1.2 This specification covers the requirements for the visual inspection, sphericity, surface finish, and dimensional tolerances for the spherical articulating metal or ceramic femoral heads of total hip joint replacements.
1.3 This specification covers the requirements for the visual inspection, surface finish, and dimensional tolerances for the spherical concave mating surface of polyethylene acetabular components; and the requirements for the visual inspection, sphericity, surface finish, and diametral clearance for the spherical concave mating surface of metal and ceramic acetabular components of total hip joint replacements.
1.4 This specification covers the requirements for the visual inspection, sphericity, surface finish, and dimensional tolerances for the spherical metal or ceramic femoral heads of partial hip joint replacements.
1.5 This specification covers the requirements for the visual inspection, sphericity, surface finish, and dimensional tolerances for the spherical articulating metal or ceramic femoral heads of bipolar femoral hip joint replacements; the requirements for the visual inspection, surface finish, and dimensional tolerances for the inner polyethylene bearing surface of bipolar heads; the requirements for the visual inspection, sphericity, surface finish, and diametral clearance of the inner metal or ceramic bearing surface of bipolar heads; and the requirements for the visual inspection, sphericity, surface finish, and dimensional tolerances of outer metal or ceramic bearing surfaces of bipolar heads of bipolar femoral hip joint replacements.
1.6 This specification covers the requirements for the visual inspection, sphericity, surface finish, and dimensional tolerances for the spherical articulating metal or ceramic femoral heads of dual mobility hip joint replacements; the requirements for the visual inspection, surface finish, and dimensional tolerances for the inner polyethylene bearing surface of dual mobility heads; the requirements for the visual inspection, sphericity, and surface finish of outer polyethylene bearing surface of dual mobility heads; and the requirements for the visual inspection, sphericity, surface finish, and diametral clearance requirements for the spherical concave mating surface of metal and ceramic acetabular components of dual mobility hip joint replacements. ... |
|
| Extent of Recognition
|
Rationale for Recognition
| This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Transition Period
| FDA recognition of ASTM F2033-20 [Rec# 11-376] will be superseded by recognition of ASTM F2033-25 [Rec# 11-437]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 11-376] until December 19, 2027. After this transition period, declarations of conformity to [Rec# 11-376] will not be accepted. |
|
Public Law, CFR Citation(s) and Procode(s)*
| Regulation Number |
Device Name |
Device Class |
Product Code |
| §888.3300 |
Prosthesis, Hip, Constrained, Metal
|
Class 3
|
KXD
|
| §888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer
|
Class 2
|
KWZ
|
| §888.3310 |
Prosthesis, Hip, Constrained, Cemented Or Uncemented, Metal/Polymer, + Additive
|
Class 2
|
PBI
|
| §888.3320 |
Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)
|
Class 3
|
JDL
|
| §888.3330 |
Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
|
Class 3
|
KWA
|
| §888.3340 |
Prosthesis, Hip, Semi-Constrained, Composite/Metal
|
Class 2
|
KMC
|
| §888.3350 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
|
Class 2
|
JDI
|
| §888.3350 |
Hip, Semi-Constrained, Cemented, Metal/Polymer + Additive, Cemented
|
Class 2
|
OQH
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
|
Class 2
|
LZO
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish
|
Class 2
|
MAY
|
| §888.3353 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
|
Class 2
|
MEH
|
| §888.3353 |
Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented
|
Class 2
|
OQI
|
| §888.3358 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
|
Class 2
|
LPH
|
| §888.3358 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
|
Class 2
|
MBL
|
| §888.3358 |
Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented
|
Class 2
|
OQG
|
| §888.3360 |
Prosthesis, Upper Femoral
|
Class 2
|
JDD
|
| §888.3360 |
Prosthesis, Hip, Femoral Component, Cemented, Metal
|
Class 2
|
JDG
|
| §888.3360 |
Prosthesis, Hip, Hemi-, Femoral, Metal
|
Class 2
|
KWL
|
| §888.3360 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented
|
Class 2
|
LWJ
|
| §888.3390 |
Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
|
Class 2
|
KWY
|
| §888.3410 |
Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer
|
Class 3
|
KXB
|
| §888.3410 |
Prosthesis, Hip, Pelvifemoral Resurfacing, Metal/Polymer, Uncemented
|
Class 3
|
OCG
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic, Cemented |
Class 3
|
LPF
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented |
Class 3
|
MRA
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing |
Class 3
|
NXT
|
| N/A |
Prosthesis, Hip, Semi-Constrained, Ceramic-On-Metal Articulation |
Class 3
|
OVO
|
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance Document for the Preparation of Premarket Notification for Ceramic Ball Hip Systems, issued January 1995.
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
| FDA Technical Contact
|
| Standards Development Organization
|
| FDA Specialty Task Group (STG)
|
| *These are provided as examples and others may be applicable. |
|
|