Device sv70 ventilator
Regulation Description Continuous ventilator.
Product CodeMNT
Regulation Number 868.5895
Device Class 2


Premarket Reviews
ManufacturerDecision
BREAS MEDICAL AB
  SUBSTANTIALLY EQUIVALENT 1
FISHER & PAYKEL HEALTHCARE LIMITED
  SUBSTANTIALLY EQUIVALENT 1
  1.  K191624  F&P Nivairo RT047 Vented Hospital Full Face Mask, ...
FISHER & PAYKEL HEALTHCARE, LTD.
  SUBSTANTIALLY EQUIVALENT 1
MAQUET CRITICAL CARE AB
  SUBSTANTIALLY EQUIVALENT 1
MEKICS CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1
NIHON KOHDEN ORANGEMED, INC.
  SUBSTANTIALLY EQUIVALENT 1
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1

MDR Year MDR Reports MDR Events
2020 4170 4170
2021 3790 3802
2022 3277 3285
2023 4565 4565
2024 4808 4809
2025 3588 3589
2026 2817 2817

Device Problems MDRs with this Device Problem Events in those MDRs
Inadequate User Interface 3900 3901
Electrical /Electronic Property Problem 2736 2737
Device Sensing Problem 1474 1474
Circuit Failure 1428 1428
Device Alarm System 1080 1080
Unexpected Shutdown 908 908
Degraded 886 894
Failure to Sense 850 850
Battery Problem 720 720
Power Problem 670 670
Gas/Air Leak 614 614
Mechanical Problem 592 592
Excessive Heating 541 541
Failure to Power Up 503 503
Complete Loss of Power 485 485
Failure to Run on Battery 459 459
No Display/Image 421 421
Failure to Deliver 356 356
Tidal Volume Fluctuations 338 338
Display Difficult to Read 297 297
Defective Alarm 295 295
Unstable 288 288
Unintended Power Up 220 220
Failure to Charge 214 214
Display or Visual Feedback Problem 213 213
Grounding Malfunction 212 212
Pressure Problem 184 184
Communication or Transmission Problem 149 149
Insufficient Flow or Under Infusion 148 148
Obstruction of Flow 144 144
Computer Software Problem 127 127
No Audible Alarm 117 117
No Flow 116 116
Insufficient Device Problem Information 109 109
Infusion or Flow Problem 106 106
Increase in Pressure 98 98
Failure to Shut Off 93 93
Volume Accuracy Problem 90 90
Noise, Audible 87 87
Improper Flow or Infusion 83 83
Adverse Event Without Identified Device or Use Problem 73 73
Material Integrity Problem 58 58
Mechanical Jam 48 48
Gas Output Problem 48 48
Audible Prompt/Feedback Problem 43 43
Decrease in Pressure 42 42
Intermittent Loss of Power 39 39
Erratic or Intermittent Display 37 37
Contamination 36 36
No Audible Prompt/Feedback 35 35

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Clinical Signs, Symptoms or Conditions 23593 23602
No Patient Involvement 1499 1499
No Known Impact Or Consequence To Patient 713 713
Insufficient Information 598 598
No Consequences Or Impact To Patient 390 390
Hypoxia 116 116
Low Oxygen Saturation 111 111
Dyspnea 48 49
Death 29 29
Unspecified Respiratory Problem 21 21
Cardiac Arrest 13 13
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available 10 10
Respiratory Arrest 9 9
Discomfort 7 19
Bradycardia 7 7
Respiratory Failure 7 7
Injury 6 6
Cyanosis 5 5
Increased Respiratory Rate 5 5
Respiratory Distress 4 4
Aspiration Pneumonitis 3 3
Abnormal Blood Gases 3 3
Hypoventilation 3 3
Anxiety 3 3
Pain 3 3
Tachycardia 3 3
No Information 3 3
Respiratory Acidosis 3 3
Bacterial Infection 3 3
Respiratory Insufficiency 3 3
Chronic Obstructive Pulmonary Disease (COPD) 3 3
Pulmonary Emphysema 2 2
Fainting 2 2
Lethargy 2 2
Loss of consciousness 2 2
Tissue Damage 2 2
Pneumonia 2 2
Angina 2 2
Apnea 2 2
Wheezing 2 2
Cardiopulmonary Arrest 2 2
Asystole 2 2
Fall 2 2
Aspiration/Inhalation 2 2
Chest Pain 2 2
Unspecified Tissue Injury 2 2
Pneumothorax 2 2
Sleepiness/Drowsiness/Somnolence 1 1
Cardiovascular Dysfunction/Insufficiency 1 1
Inflammation 1 1

Recalls
Manufacturer Recall Class Date Posted
1 DRE Medical Group Inc II Jun-29-2023
2 Philips Respironics, Inc. I Jul-02-2025
3 Philips Respironics, Inc. I May-24-2024
4 Philips Respironics, Inc. I Sep-28-2023
5 Philips Respironics, Inc. II May-21-2021
6 ResMed Ltd. I Feb-07-2020
7 Respironics California, LLC I Apr-18-2022
8 Respironics California, LLC I Feb-25-2022
9 Respironics California, LLC I Jul-22-2021
10 Respironics California, LLC II Feb-19-2021
11 Respironics California, LLC II Jan-22-2021
12 Respironics California, LLC II Jan-22-2021
13 Respironics California, LLC II Mar-31-2020
14 Respironics California, LLC II Mar-02-2020
-
-