Device system, ablation, microwave and accessories
Regulation Description Electrosurgical cutting and coagulation device and accessories.
Product CodeNEY
Regulation Number 878.4400
Device Class 2


Premarket Reviews
ManufacturerDecision
ANGIODYNAMICS, INC.
  SUBSTANTIALLY EQUIVALENT 3
BETTERS (SUZHOU)MEDICAL CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1
BETTERS(SUZHOU) MEDICAL CO., LTD
  SUBSTANTIALLY EQUIVALENT 1
CANYON MEDICAL INC.
  SUBSTANTIALLY EQUIVALENT 2
COVIDIEN LLC
  SUBSTANTIALLY EQUIVALENT 2
COVIDIEN, LLC
  SUBSTANTIALLY EQUIVALENT 1
CREO MEDICAL LTD.
  SUBSTANTIALLY EQUIVALENT 4
EMBLATION LIMITED
  SUBSTANTIALLY EQUIVALENT 1
EMBLATION LTD.
  SUBSTANTIALLY EQUIVALENT 1
GALIL MEDICAL INC.
  SUBSTANTIALLY EQUIVALENT 1
HYGEA MEDICAL TECHNOLOGY CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1
MEDTRONIC, INC
  SUBSTANTIALLY EQUIVALENT 1
MEDWAVES, INC.
  SUBSTANTIALLY EQUIVALENT 1
MIMA-PRO SCIENTIFIC INC.
  SUBSTANTIALLY EQUIVALENT 1
NANJING ECO MICROWAVE SYSTEM CO., LTD
  SUBSTANTIALLY EQUIVALENT 1
  1.  K201262  Microwave Therapeutic System
NANJING ECO MICROWAVE SYSTEM CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1
NEUWAVE MEDICAL INC.
  SUBSTANTIALLY EQUIVALENT 1
NEUWAVE MEDICAL, INC.
  SUBSTANTIALLY EQUIVALENT 2
VARIAN MEDICAL SYSTEMS, INC.
  SUBSTANTIALLY EQUIVALENT 2

MDR Year MDR Reports MDR Events
2020 128 128
2021 217 217
2022 142 142
2023 107 107
2024 141 141
2025 25 25

Device Problems MDRs with this Device Problem Events in those MDRs
Adverse Event Without Identified Device or Use Problem 258 258
Detachment of Device or Device Component 163 163
Appropriate Term/Code Not Available 126 126
Break 96 96
Temperature Problem 30 30
Leak/Splash 21 21
Device Alarm System 17 17
No Flow 16 16
Energy Output Problem 14 14
Material Twisted/Bent 12 12
Device Displays Incorrect Message 11 11
Fracture 11 11
Activation Problem 11 11
Activation, Positioning or Separation Problem 9 9
Insufficient Information 9 9
Failure to Fire 8 8
Use of Device Problem 8 8
Gas/Air Leak 8 8
Connection Problem 6 6
Material Integrity Problem 6 6
Intermittent Energy Output 6 6
Manufacturing, Packaging or Shipping Problem 6 6
Excessive Heating 5 5
Thermal Decomposition of Device 5 5
Difficult or Delayed Activation 5 5
Material Separation 4 4
Defective Device 4 4
Failure to Read Input Signal 3 3
Insufficient Heating 3 3
Difficult to Remove 3 3
Therapeutic or Diagnostic Output Failure 3 3
Material Discolored 3 3
Biocompatibility 3 3
Visual Prompts will not Clear 3 3
Output Problem 3 3
Protective Measures Problem 3 3
Fluid/Blood Leak 3 3
Failure to Deliver Energy 3 3
Firing Problem 2 2
Peeled/Delaminated 2 2
Crack 2 2
Failure to Power Up 2 2
Flaked 2 2
Material Fragmentation 2 2
Melted 2 2
Pumping Problem 2 2
Continuous Firing 2 2
Failure to Shut Off 2 2
Delivered as Unsterile Product 2 2
Smoking 2 2

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Clinical Signs, Symptoms or Conditions 217 217
Device Embedded In Tissue or Plaque 128 128
Insufficient Information 76 76
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available 67 67
Foreign Body In Patient 65 65
Pneumothorax 52 52
Hemorrhage/Bleeding 43 43
Pain 32 32
Pleural Effusion 31 31
Abscess 28 28
Hematoma 25 25
No Code Available 24 24
No Consequences Or Impact To Patient 21 21
Burn(s) 20 20
Unspecified Tissue Injury 15 15
Obstruction/Occlusion 14 14
Fistula 12 12
Thrombosis/Thrombus 12 12
High Blood Pressure/ Hypertension 12 12
Abdominal Pain 11 11
Unspecified Infection 11 11
Fever 11 11
No Known Impact Or Consequence To Patient 10 10
Hypersensitivity/Allergic reaction 8 8
Full thickness (Third Degree) Burn 7 7
Unintended Radiation Exposure 7 7
Hemothorax 6 6
Bacterial Infection 6 6
Radiation Exposure, Unintended 6 6
Hemoptysis 6 6
Chest Pain 6 6
Embolism/Embolus 5 5
Sepsis 5 5
No Information 5 5
Pneumonia 5 5
Unspecified Hepatic or Biliary Problem 5 5
Cancer 5 5
Peritonitis 4 4
Ascites 4 4
Urinary Retention 4 4
Fatigue 4 4
Bone Fracture(s) 4 4
Vomiting 4 4
Death 4 4
Anemia 3 3
Pulmonary Embolism 3 3
Nerve Damage 3 3
Nausea 3 3
Perforation 3 3
Renal Failure 3 3

Recalls
Manufacturer Recall Class Date Posted
1 Angiodynamics, Inc. II Sep-08-2021
2 Covidien Llc II Nov-05-2021
3 Covidien, LLC II Jan-12-2023
-
-