• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

TPLC - Total Product Life Cycle

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search show TPLC since Back to Search Results
Device heater, breathing system w/wo controller (not humidifier or nebulizer
Regulation Description Breathing system heater.
Product CodeBZE
Regulation Number 868.5270
Device Class 2


Premarket Reviews
ManufacturerDecision
DRAEGERWERK AG & CO KGAA
  SUBSTANTIALLY EQUIVALENT 1
  1.  K222822  VentStar Helix dual heated (N) Plus (MP02650), Ven ...
EXCELERON MEDICAL
  SUBSTANTIALLY EQUIVALENT 1
GUANGDONG EDA TECHNOLOGY CO., LTD.
  SUBSTANTIALLY EQUIVALENT 1
MEDLINE INDUSTRIES, LP
  SUBSTANTIALLY EQUIVALENT 1
SUNSET HEALTHCARE SOLUTIONS, INC.
  SUBSTANTIALLY EQUIVALENT 1

MDR Year MDR Reports MDR Events
2020 123 123
2021 124 124
2022 152 154
2023 125 129
2024 172 175
2025 195 195
2026 14 14

Device Problems MDRs with this Device Problem Events in those MDRs
Leak/Splash 318 321
Gas/Air Leak 296 297
Crack 119 120
Break 52 53
Fluid/Blood Leak 46 48
Material Split, Cut or Torn 41 41
Therapeutic or Diagnostic Output Failure 40 40
Insufficient Heating 30 30
Disconnection 29 29
Failure of Device to Self-Test 28 28
Connection Problem 23 23
Material Puncture/Hole 15 15
Loose or Intermittent Connection 14 14
Mechanical Problem 13 13
Incorrect, Inadequate or Imprecise Result or Readings 12 12
Detachment of Device or Device Component 11 11
Degraded 11 11
Fracture 7 7
Use of Device Problem 6 7
Material Deformation 5 5
Misconnection 5 6
Moisture or Humidity Problem 5 5
Melted 5 5
Component Missing 5 5
Manufacturing, Packaging or Shipping Problem 5 5
Material Integrity Problem 4 4
Infusion or Flow Problem 4 4
Improper or Incorrect Procedure or Method 3 3
Obstruction of Flow 3 3
Corroded 3 3
Material Perforation 3 3
Failure to Disconnect 3 3
Output Problem 3 3
Misassembled 3 3
Fire 3 3
Misassembly by Users 3 3
Fitting Problem 3 3
Overfill 2 2
Misassembled During Installation 2 2
Insufficient Information 2 2
Material Twisted/Bent 2 2
Appropriate Term/Code Not Available 2 2
Ambient Temperature Problem 2 2
Protective Measures Problem 2 2
Overheating of Device 1 1
Device Displays Incorrect Message 1 1
Deformation Due to Compressive Stress 1 1
No Flow 1 1
Loss of or Failure to Bond 1 1
Noise, Audible 1 1

Patient Problems MDRs with this Patient Problem Events in those MDRs
No Clinical Signs, Symptoms or Conditions 778 786
No Patient Involvement 50 50
No Known Impact Or Consequence To Patient 28 28
Low Oxygen Saturation 22 23
Insufficient Information 12 12
No Consequences Or Impact To Patient 11 11
Hypoxia 4 4
Aspiration/Inhalation 3 3
Burning Sensation 3 3
Pneumothorax 2 2
Dizziness 2 2
Bradycardia 2 2
Headache 2 2
Hypoventilation 1 1
Death 1 1
Respiratory Arrest 1 1
Abnormal Blood Gases 1 1
Discomfort 1 1
Dyspnea 1 1
Laceration(s) 1 1
Increased Respiratory Rate 1 1
Tachycardia 1 1
Unspecified Respiratory Problem 1 1
Irritability 1 1
Unspecified Vascular Problem 1 1
Respiratory Failure 1 1
Increased Peak Expiratory Flow rate 1 1

Recalls
Manufacturer Recall Class Date Posted
1 Vyaire Medical I May-09-2025
-
-