Device
pulse-generator, pacemaker, external
Regulation Description
External pacemaker pulse generator.
Product Code
DTE
Regulation Number
870.3600
Device Class
2
Premarket Reviews
Manufacturer
Decision
SHENZHEN LIFETECH CARDIO MEDICAL ELECTRONICS CO., LTD.
SUBSTANTIALLY EQUIVALENT
1
SOLO PACE, INC.
SUBSTANTIALLY EQUIVALENT
1
1. K241781
Solo Pace Control
MDR Year
MDR Reports
MDR Events
2021
899
899
2022
833
835
2023
650
651
2024
620
620
2025
643
643
2026
131
131
Device Problems
MDRs with this Device Problem
Events in those MDRs
Connection Problem
611
611
No Display/Image
609
609
Mechanical Problem
591
591
Calibration Problem
376
376
Display or Visual Feedback Problem
346
346
Output Problem
330
330
Device Sensing Problem
237
237
Unexpected Shutdown
216
216
No Device Output
202
202
Use of Device Problem
199
199
Failure to Power Up
197
197
Break
165
166
Material Integrity Problem
135
135
Corroded
115
115
Battery Problem
96
96
Component Missing
87
87
Contamination
81
81
Electrical /Electronic Property Problem
74
74
Failure to Capture
66
66
Crack
55
55
Loose or Intermittent Connection
43
43
Capturing Problem
35
35
Power Problem
33
33
Pacing Problem
24
24
Under-Sensing
20
20
Pacing Intermittently
20
20
Output below Specifications
16
16
Adverse Event Without Identified Device or Use Problem
16
17
Moisture Damage
16
16
Failure to Sense
13
13
Intermittent Capture
12
13
Over-Sensing
9
9
Premature Discharge of Battery
8
8
Output above Specifications
8
8
Failure to Shut Off
8
8
Failure of Device to Self-Test
6
6
Sensing Intermittently
6
6
Failure to Calibrate
6
6
Inadequate User Interface
5
5
Key or Button Unresponsive/not Working
4
4
Operating System Becomes Nonfunctional
4
4
Blocked Connection
4
4
Circuit Failure
4
4
No Pacing
3
3
Application Program Freezes, Becomes Nonfunctional
3
3
Device Contamination with Chemical or Other Material
3
3
Insufficient Information
3
3
Device Difficult to Program or Calibrate
3
3
Impedance Problem
3
3
Contamination /Decontamination Problem
3
3
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
No Clinical Signs, Symptoms or Conditions
3606
3607
Cardiac Arrest
40
40
Ventricular Fibrillation
31
31
Asystole
31
31
Arrhythmia
24
25
Tachycardia
17
17
Bradycardia
13
13
Heart Block
8
8
Insufficient Information
8
8
Loss of consciousness
7
7
Superficial (First Degree) Burn
6
6
Low Blood Pressure/ Hypotension
5
6
Atrial Flutter
4
4
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available
3
4
Erythema
3
3
Non specific EKG/ECG Changes
2
2
Syncope/Fainting
2
2
Diaphoresis
2
2
Low Oxygen Saturation
2
2
Dizziness
2
2
Skin Inflammation/ Irritation
2
2
Aspiration/Inhalation
2
2
Abrasion
2
2
Multiple Organ Failure
2
2
Pericardial Effusion
2
3
Blister
2
2
Unspecified Heart Problem
2
2
Discomfort
2
2
Pneumonia
2
2
Bruise/Contusion
1
1
Respiratory Failure
1
1
Hemorrhage/Bleeding
1
1
Ectopic Heartbeat
1
1
Hypoxia
1
1
Swelling/ Edema
1
1
Sudden Cardiac Death
1
1
Stenosis
1
1
Shock
1
1
Pulmonary Edema
1
1
Dyspnea
1
1
Septic Shock
1
1
Cardiac Perforation
1
1
Pallor
1
1
Burn(s)
1
1
Angina
1
1
Cardiac Tamponade
1
1
Skin Discoloration
1
1
Device Overstimulation of Tissue
1
1
Muscle Weakness
1
1
Perforation
1
2
-
-