Device
catheter, peripheral, atherectomy
Product Code
MCW
Regulation Number
870.4875
Device Class
2
Premarket Reviews
Manufacturer
Decision
AVINGER, INC.
SUBSTANTIALLY EQUIVALENT
5
BOSTON SCIENTIFIC CORPORATION
SUBSTANTIALLY EQUIVALENT
1
C.R. BARD, INC.
SUBSTANTIALLY EQUIVALENT
1
CARDIO FLOW INC.
SUBSTANTIALLY EQUIVALENT
1
CARDIO FLOW, INC.
SUBSTANTIALLY EQUIVALENT
1
CARDIOVASCUALR SYSTEMS INC.
SUBSTANTIALLY EQUIVALENT
1
CARDIOVASCULAR SYSTEMS INC.
SUBSTANTIALLY EQUIVALENT
6
CARDIOVASCULAR SYSTEMS, INC.
SUBSTANTIALLY EQUIVALENT
1
EXIMO MEDICAL LTD.
SUBSTANTIALLY EQUIVALENT
5
EXIMO MEDICAL, LTD.
SUBSTANTIALLY EQUIVALENT
1
MEDTRONIC VASCULAR, INC.
SUBSTANTIALLY EQUIVALENT
1
MEDTRONIC, INC
SUBSTANTIALLY EQUIVALENT
1
REX MEDICAL, L.P.
SUBSTANTIALLY EQUIVALENT
2
SPECTRANETICS INC.
SUBSTANTIALLY EQUIVALENT
1
SPECTRANETICS, INC.
SUBSTANTIALLY EQUIVALENT
4
STRAUB MEDICAL AG
SUBSTANTIALLY EQUIVALENT
1
VOLCANO ATHEROMED INC
SUBSTANTIALLY EQUIVALENT
1
VOLCANO ATHEROMED INC.
SUBSTANTIALLY EQUIVALENT
2
VOLCANO ATHEROMED, INC.
SUBSTANTIALLY EQUIVALENT
1
MDR Year
MDR Reports
MDR Events
2017
415
415
2018
537
537
2019
618
618
2020
631
753
2021
734
2930
2022
566
566
2023
575
575
2024
145
145
Device Problems
MDRs with this Device Problem
Events in those MDRs
Detachment of Device or Device Component
674
918
Entrapment of Device
671
671
Break
611
1221
Adverse Event Without Identified Device or Use Problem
580
1556
Material Separation
372
372
Material Deformation
322
444
Difficult to Remove
245
245
Mechanical Problem
225
225
Physical Resistance/Sticking
213
213
Mechanical Jam
174
174
Suction Problem
170
170
Fracture
138
138
Device Damaged by Another Device
120
120
Difficult to Advance
97
97
Noise, Audible
80
80
Leak/Splash
68
68
Use of Device Problem
67
67
Device Operates Differently Than Expected
53
53
Retraction Problem
51
51
Failure to Advance
44
44
Aspiration Issue
43
43
Positioning Problem
42
42
Material Integrity Problem
32
32
Device-Device Incompatibility
32
32
Defective Device
31
31
Detachment Of Device Component
30
30
Device Markings/Labelling Problem
28
28
Material Twisted/Bent
26
26
Unexpected Shutdown
26
26
Material Split, Cut or Torn
25
25
Fluid/Blood Leak
25
25
Infusion or Flow Problem
24
24
Peeled/Delaminated
21
21
Overheating of Device
20
20
Material Puncture/Hole
20
20
Intermittent Loss of Power
17
17
Deformation Due to Compressive Stress
16
16
Melted
15
15
Device Difficult to Setup or Prepare
14
14
Off-Label Use
14
14
Display or Visual Feedback Problem
13
13
Device Displays Incorrect Message
13
13
Material Rupture
13
13
Device Remains Activated
13
13
Unintended System Motion
12
12
No Apparent Adverse Event
12
12
Device Contamination with Chemical or Other Material
12
12
Device Dislodged or Dislocated
11
11
Insufficient Information
11
11
Stretched
11
11
Improper or Incorrect Procedure or Method
11
11
Difficult or Delayed Activation
11
11
Device Operational Issue
10
10
Hole In Material
10
10
Unintended Movement
10
10
Activation Problem
10
10
Compatibility Problem
10
10
Difficult to Open or Close
9
9
Contamination
9
9
Crack
9
253
Failure to Prime
9
9
Material Fragmentation
8
8
Output Problem
8
8
Missing Information
8
8
Therapeutic or Diagnostic Output Failure
8
8
No Flow
7
7
Device Handling Problem
7
7
Material Frayed
7
7
Defective Component
7
7
Material Perforation
7
7
Difficult to Insert
6
6
Appropriate Term/Code Not Available
6
6
Air/Gas in Device
6
6
Inaccurate Information
6
6
Torn Material
5
5
Migration
5
5
Material Protrusion/Extrusion
5
5
Activation, Positioning or Separation Problem
5
5
Loss of or Failure to Bond
5
5
Failure to Power Up
5
5
Obstruction of Flow
5
5
Connection Problem
5
5
Device Misassembled During Manufacturing /Shipping
4
4
Contamination /Decontamination Problem
4
4
Failure to Calibrate
4
126
Device Contamination with Body Fluid
4
4
Insufficient Flow or Under Infusion
4
4
Kinked
4
4
Migration or Expulsion of Device
4
4
Device Packaging Compromised
4
4
Failure to Shut Off
4
4
Activation Failure
4
4
Sharp Edges
4
4
Wrong Label
3
3
Temperature Problem
3
3
Packaging Problem
3
3
Power Problem
3
3
Device Contaminated During Manufacture or Shipping
3
3
Difficult to Flush
3
3
Complete Blockage
3
3
Patient Problems
MDRs with this Patient Problem
Events in those MDRs
No Clinical Signs, Symptoms or Conditions
1437
2047
No Consequences Or Impact To Patient
1116
1116
Foreign Body In Patient
304
304
No Known Impact Or Consequence To Patient
204
326
Vascular Dissection
191
557
Perforation of Vessels
159
281
Device Embedded In Tissue or Plaque
151
151
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available
114
114
Embolism/Embolus
114
358
Embolism
110
232
Perforation
75
75
Insufficient Information
72
72
Patient Problem/Medical Problem
65
65
No Code Available
51
51
Unintended Radiation Exposure
49
903
Pain
49
49
Radiation Exposure, Unintended
47
169
Vessel Or Plaque, Device Embedded In
42
42
Thrombosis/Thrombus
41
41
Injury
39
39
Reocclusion
37
37
Occlusion
29
29
Vascular System (Circulation), Impaired
29
29
Pseudoaneurysm
28
28
Stenosis
28
28
Obstruction/Occlusion
27
27
Iatrogenic Source
26
26
Hematoma
24
24
Thrombus
22
22
Intimal Dissection
19
19
Unspecified Tissue Injury
19
19
No Patient Involvement
17
17
Aneurysm
17
17
Death
17
17
Restenosis
16
16
Hemorrhage/Bleeding
13
13
Myocardial Infarction
13
13
Thromboembolism
12
12
Thrombosis
11
11
Fistula
10
10
Embolus
10
10
Extravasation
10
10
Low Blood Pressure/ Hypotension
9
9
Ischemia
9
9
Rupture
9
9
Sepsis
8
8
Vasoconstriction
7
7
Renal Failure
6
6
Cardiac Arrest
6
6
Peripheral Vascular Disease
5
5
Swelling/ Edema
5
5
Intraoperative Pain
5
5
Blood Loss
4
4
Foreign Body Embolism
4
4
Tissue Damage
4
4
Fever
4
4
Stroke/CVA
4
4
Nausea
3
3
Calcium Deposits/Calcification
3
3
Hypersensitivity/Allergic reaction
3
3
No Information
3
3
Laceration(s)
2
2
Muscle Spasm(s)
2
2
Chest Pain
2
2
Dyspnea
2
2
Ecchymosis
2
2
Pulmonary Embolism
2
2
Arrhythmia
2
2
Burning Sensation
2
2
Tachycardia
2
2
Deformity/ Disfigurement
2
2
Atrial Perforation
2
2
Discomfort
2
2
Complaint, Ill-Defined
1
1
Respiratory Acidosis
1
1
Respiratory Failure
1
1
Shaking/Tremors
1
1
Skin Tears
1
1
Claudication
1
1
Disability
1
1
Loss of consciousness
1
1
Cardiac Tamponade
1
1
Anxiety
1
1
Pneumonia
1
1
Great Vessel Perforation
1
1
Heart Failure
1
1
Vomiting
1
1
Bradycardia
1
1
Bruise/Contusion
1
1
Burn(s)
1
1
Cardiopulmonary Arrest
1
1
Exsanguination
1
1
Failure of Implant
1
1
Unspecified Infection
1
1
High Blood Pressure/ Hypertension
1
1
Hyperplasia
1
1
Foreign Body Reaction
1
1
Bone Fracture(s)
1
1
Gangrene
1
1
Missing Value Reason
1
1
Recalls
Manufacturer
Recall Class
Date Posted
1
Boston Scientific Corporation
II
Apr-10-2017
2
Cardiovascular Systems Inc
II
Aug-28-2021
3
Cardiovascular Systems Inc
II
Oct-11-2018
4
Cardiovascular Systems Inc
II
Apr-10-2017
5
Cardiovascular Systems Inc
II
Apr-10-2017
6
Cardiovascular Systems Inc
II
Feb-22-2017
7
Micro Therapeutics Inc, Dba Ev3 Neurovascular
II
Sep-11-2018
8
Spectranetics Corporation
II
Dec-08-2023
9
Spectranetics Corporation
II
Jul-14-2017
10
ev3 Inc.
I
Mar-07-2022
11
ev3 Inc.
I
Jan-19-2022
-
-