Device growing rod system- magnetic actuation
Regulation Description Thoracolumbosacral pedicle screw system.
Definition Stabilization or correction of spinal deformities without the use of fusion, using non-invasive methods to adjust the spinal rod.
Product CodePGN
Regulation Number 888.3070
Device Class 2


Premarket Reviews
ManufacturerDecision
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
  SUBSTANTIALLY EQUIVALENT 1
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
  SUBSTANTIALLY EQUIVALENT 1

MDR Year MDR Reports MDR Events
2020 339 339
2021 204 204
2022 120 120
2023 189 189
2024 30 31
2025 4 4

Device Problems MDRs with this Device Problem Events in those MDRs
Defective Device 178 178
Mechanical Jam 142 142
Corroded 130 130
Mechanical Problem 128 128
Break 112 112
Failure to Advance 95 96
Migration or Expulsion of Device 55 55
Adverse Event Without Identified Device or Use Problem 53 53
Fracture 50 50
Migration 42 42
Physical Resistance/Sticking 39 39
Naturally Worn 34 34
Degraded 28 28
Scratched Material 28 28
Structural Problem 27 27
Detachment of Device or Device Component 24 24
Defective Component 24 24
Insufficient Information 17 17
No Apparent Adverse Event 15 15
Unstable 15 15
Retraction Problem 7 7
Material Protrusion/Extrusion 5 5
Appropriate Term/Code Not Available 5 5
Material Erosion 5 5
Material Twisted/Bent 3 3
Noise, Audible 3 3
Material Discolored 3 3
Dent in Material 2 2
Loosening of Implant Not Related to Bone-Ingrowth 2 2
Failure to Seal 1 1
Malposition of Device 1 1
Activation, Positioning or Separation Problem 1 1
Unintended Movement 1 1
Particulates 1 1
Fluid/Blood Leak 1 1
Component Missing 1 1
Electromagnetic Interference 1 1
Therapeutic or Diagnostic Output Failure 1 1
Crack 1 1
Material Deformation 1 1
Material Fragmentation 1 1

Patient Problems MDRs with this Patient Problem Events in those MDRs
Failure of Implant 694 695
No Clinical Signs, Symptoms or Conditions 74 74
Insufficient Information 48 49
Metal Related Pathology 28 28
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available 25 25
Unspecified Infection 25 25
Pain 14 14
Skin Discoloration 5 5
Fever 4 4
Inflammation 4 4
Purulent Discharge 4 4
Drug Resistant Bacterial Infection 3 3
Wound Dehiscence 3 3
Implant Pain 2 2
Bacterial Infection 2 2
Deformity/ Disfigurement 2 2
Decreased Appetite 2 2
Spinal Cord Injury 2 2
Joint Laxity 2 2
Burning Sensation 2 2
No Information 2 2
Localized Skin Lesion 2 2
Necrosis 2 2
Ossification 2 2
Hemorrhage/Bleeding 2 2
No Code Available 2 2
Cerebral Ventriculomeglia 1 1
Infiltration into Tissue 1 1
Fall 1 1
Burn(s) 1 1
Headache 1 1
Post Operative Wound Infection 1 1
Local Reaction 1 1
Foreign Body Reaction 1 1
Skin Erosion 1 1
Sepsis 1 1
Fluid Discharge 1 1
Seizures 1 1
No Known Impact Or Consequence To Patient 1 1

Recalls
Manufacturer Recall Class Date Posted
1 Globus Medical, Inc. II Apr-23-2025
2 NuVasive Inc II May-05-2020
3 NuVasive Specialized Orthopedics, Inc. II Dec-14-2022
4 NuVasive Specialized Orthopedics, Inc. II Jan-26-2022
-
-