[Code of Federal Regulations]
[Title 21, Volume 1]
[CITE: 21CFR4]





TITLE 21--FOOD AND DRUGS
CHAPTER I--FOOD AND DRUG ADMINISTRATION
DEPARTMENT OF HEALTH AND HUMAN SERVICES
SUBCHAPTER A - GENERAL
 
PART 4REGULATION OF COMBINATION PRODUCTS
 

Subpart B - Postmarketing Safety Reporting for Combination Products

Sec. 4.100 What is the scope of this subpart?

(a) This subpart identifies postmarketing safety reporting requirements for combination product applicants and constituent part applicants.

(b) This subpart does not apply to investigational combination products, combination products that have not received marketing authorization, or to persons other than combination product applicants and constituent part applicants.

(c) This subpart supplements and does not supersede other provisions of this chapter, including the provisions in parts 314, 600, 606, 803, and 806 of this chapter, unless a regulation explicitly provides otherwise.

Sec. 4.101 How does the FDA define key terms and phrases in this subpart?

Abbreviated new drug application (ANDA) has the same meaning given the term "abbreviated application" in § 314.3(b) of this chapter.

Agency or we means Food and Drug Administration.

Applicant means, for the purposes of this subpart, a person holding an application under which a combination product or constituent part of a combination product has received marketing authorization (such as approval, licensure, or clearance). For the purposes of this subpart, applicant is used interchangeably with the term "you."

Application means, for purposes of this subpart, a BLA, an NDA, an ANDA, or a device application, including all amendments and supplements to them.

Biological product has the meaning given the term in section 351 of the Public Health Service Act (42 U.S.C. 262).

Biological product deviation report (BPDR) is a report as described in §§ 600.14 and 606.171 of this chapter.

Biologics license application (BLA) has the meaning given the term in section 351 of the Public Health Service Act (42 U.S.C. 262) and § 601.2 of this chapter.

Combination product has the meaning given the term in § 3.2(e) of this chapter.

Combination product applicant means an applicant that holds the application(s) for a combination product.

Constituent part has the meaning given the term in § 4.2.

Constituent part applicant means the applicant for a constituent part of a combination product the constituent parts of which are marketed under applications held by different applicants.

Correction or removal report is a report as described in § 806.10 of this chapter.

De novo classification request is a submission requesting de novo classification under section 513(f)(2) of the Federal Food, Drug, and Cosmetic Act.

Device has the meaning given the term in section 201(h) of the Federal Food, Drug, and Cosmetic Act.

Device application means a PMA, PDP, premarket notification submission, de novo classification request, or HDE.

Drug has the meaning given the term in section 201(g)(1) of the Federal Food, Drug, and Cosmetic Act.

Field alert report is a report as described in § 314.81 of this chapter.

Fifteen-day report is a report required to be submitted within 15 days as described in § 314.80 of this chapter or § 600.80 of this chapter, as well as followup reports to such a report.

Five-day report is a report as described in §§ 803.3 and 803.53 of this chapter, as well as supplemental or followup reports to such a report as described in § 803.56 of this chapter.

Humanitarian device exemption (HDE) has the meaning given the term in § 814.3 of this chapter.

Malfunction report is a report as described in § 803.50 of this chapter as well as supplemental or followup reports to such a report as described in § 803.56 of this chapter.

New drug application (NDA) has the meaning given the term "application" in § 314.3(b) of this chapter.

Premarket approval application (PMA) has the meaning given the term in § 814.3 of this chapter.

Premarket notification submission is a submission as described in § 807.87 of this chapter.

Product Development Protocol (PDP) is a submission as set forth in section 515(f) of the Federal Food, Drug, and Cosmetic Act.

Sec. 4.102 What reports must you submit to FDA for your combination product or constituent part?

(a) In general. If you are a constituent part applicant, the reporting requirements applicable to you that are identified in this section apply to your constituent part, and if you are a combination product applicant, the reporting requirements applicable to you that are identified in this section apply to your combination product as a whole.

(b) Reporting requirements applicable to both combination product applicants and constituent part applicants. If you are a combination product applicant or constituent part applicant, you must comply with the reporting requirements identified in paragraphs (b)(1), (b)(2), or (b)(3) of this section for your product based on its application type. If you are a combination product applicant, you are required to submit a report as specified in this paragraph unless you have already submitted a report in accordance with paragraph (c) of this section for the same event that: Includes the information required under the applicable regulations identified in this paragraph, is required to be submitted in the same manner under § 4.104, and meets the deadlines under the applicable regulations identified in this paragraph.

(1) If your combination product or device constituent part received marketing authorization under a device application, you must comply with the requirements for postmarketing safety reporting described in parts 803 and 806 of this chapter with respect to your product.

(2) If your combination product or drug constituent part received marketing authorization under an NDA or ANDA, you must comply with the requirements for postmarketing safety reporting described in part 314 of this chapter with respect to your product.

(3) If your combination product or biological product constituent part received marketing authorization under a BLA, you must comply with the requirements for postmarketing safety reporting described in parts 600 and 606 of this chapter with respect to your product.

(c) Reporting requirements applicable only to combination product applicants. If you are a combination product applicant, in addition to compliance with paragraph (a) of this section, you must also comply with the reporting requirements identified under this paragraph as applicable to your product based on its constituent parts. If you are a combination product applicant, you are required to submit a report as specified in this paragraph unless you have already submitted a report in accordance with paragraph (b) of this section for the same event that: Includes the information required under the applicable regulations for the report identified in this paragraph; is required to be submitted in the same manner under § 4.104 of this chapter; and, unless otherwise specified in this paragraph, meets the deadlines under the applicable regulations for the report identified in this paragraph.

(1) If your combination product contains a device constituent part, you must submit:

(i) Five-day reports;

(ii) Malfunction reports; and

(iii) Correction or removal reports, and maintain records as described in § 806.20 of this chapter for corrections and removals not required to be reported.

(2) If your combination product contains a drug constituent part, you must submit:

(i) Field alert reports; and

(ii) Fifteen-day reports as described in § 314.80 of this chapter, which must be submitted within 30 calendar days instead of 15 calendar days if your combination product received marketing authorization under a device application.

(3) If your combination product contains a biological product constituent part, you must submit:

(i) Biological product deviation reports; and

(ii) Fifteen-day reports as described in § 600.80 of this chapter, which must be submitted within 30 calendar days instead of 15 calendar days if your combination product received marketing authorization under a device application.

(d) Other reporting requirements for combination product applicants. (1) If you are the combination product applicant for a combination product that contains a device constituent part and that received marketing authorization under an NDA, ANDA, or BLA, in addition to the information otherwise required in the periodic safety reports you submit under § 314.80 or § 600.80 of this chapter, your periodic safety reports must also include a summary and analysis of the reports identified in paragraphs (c)(1)(i) and (ii) of this section that were submitted during the report interval.

(2) If you are the combination product applicant for a combination product that received marketing authorization under a device application, in addition to the reports required under paragraphs (b) and (c) of this section, you must submit reports regarding postmarketing safety events if notified by the Agency in writing that the Agency requires additional information. We will specify what safety information is needed and will require such information if we determine that protection of the public health requires additional or clarifying safety information for the combination product. In any request under this section, we will state the reason or purpose for the safety information request, specify the due date for submitting the information, and clearly identify the reported event(s) related to our request.

Sec. 4.103 What information must you share with other constituent part applicants for the combination product?

(a) When you receive information regarding an event that involves a death or serious injury as described in § 803.3 of this chapter, or an adverse experience as described in § 314.80(a) of this chapter or § 600.80(a) of this chapter, associated with the use of the combination product, you must provide the information to the other constituent part applicant(s) for the combination product no later than 5 calendar days of your receipt of the information.

(b) With regard to information you must provide to the other constituent part applicant(s) for the combination product, you must maintain records that include:

(1) A copy of the information you provided,

(2) The date the information was received by you,

(3) The date the information was provided to the other constituent part applicant(s), and

(4) The name and address of the other constituent part applicant(s) to whom you provided the information.

Sec. 4.104 How and where must you submit postmarketing safety reports for your combination product or constituent part?

(a) If you are a constituent part applicant, you must submit postmarketing safety reports in accordance with the regulations identified in § 4.102(b) that are applicable to your product based on its application type.

(b) If you are a combination product applicant, you must submit postmarketing safety reports required under § 4.102 in the manner specified in the regulation applicable to the type of report, with the following exceptions:

(1) You must submit the postmarketing safety reports identified in § 4.102(c)(1)(i) and (ii) in accordance with § 314.80(g) of this chapter if your combination product received marketing authorization under an NDA or ANDA or in accordance with § 600.80(h) of this chapter if your combination product received marketing authorization under a BLA.

(2) You must submit the postmarketing safety reports identified in § 4.102(c)(2)(ii) and (c)(3)(ii) in accordance with § 803.12(a) of this chapter if your combination product received marketing authorization under a device application.

Sec. 4.105 What are the postmarketing safety reporting recordkeeping requirements for your combination product or constituent part?

(a) If you are a constituent part applicant:

(1) You must maintain records in accordance with the recordkeeping requirements in the applicable regulation(s) described in § 4.102(b).

(2) You must maintain records required under § 4.103(b) for the longest time period required for records under the postmarketing safety reporting regulations applicable to your product under § 4.102(b).

(b) If you are a combination product applicant, you must maintain records in accordance with the longest time period required for records under the regulations applicable to your product under § 4.102.

Authority: 21 U.S.C. 321, 331, 351, 352, 353, 355, 360, 360b-360f, 360h-360j, 360l, 360hh-360ss, 360aaa-360bbb, 371(a), 372-374, 379e, 381, 383, 394; 42 U.S.C. 216, 262, 263a, 264, 271.
Source: 78 FR 4321, Jan. 22, 2013, unless otherwise noted.

-
-