| Trade Name | NeuRx Diaphragm Pacing System |
| Classification Name | diaphragmatic/phrenic nerve laparoscopically-implanted stimulator |
| Generic Name | diaphragmatic/phrenic nerve laparoscopically-implanted stimulator |
| Applicant |
| SYNAPSE BIOMEDICAL |
| 300 artino street |
| oberlin, OH 44074 |
|
| HDE Number | H070003 |
| Supplement Number | S015 |
| Date Received | 04/03/2019 |
| Decision Date | 11/12/2019 |
| Product Code | |
| Advisory Committee |
Anesthesiology |
| Supplement Type | normal 75 day track |
| Supplement Reason | change design/components/specifications/material |
| Expedited Review Granted? | No |
| Combination Product | No |
Approval Order Statement Approval for design updates to the external pulse generator (epg) component of the neurx® diaphragm pacing system. These updates include the following:• primary battery options expanded from lithium metal batteries to include c-size alkaline battery. • primary battery lifetime extended by twenty percent under nominal stimulation settings. • ingress protection rating improvement (ipx4 to ip24). • programming connector accessibility improved to eliminate the need to remove primary battery for access. • twenty-two percent reduction in volumetric size. • prioritized audible alarms for error conditions. • unique alarm patterns for high, medium, and low priority error conditions. • alarm pattern repetition rate increased. • prioritized display messaging for error conditions. |