| Trade Name | Reclaim DBS Therapy for Obsessive Compulsive Disorder (OCD) System |
| Classification Name | deep brain stimulator for obsessive compulsive disorder (ocd) |
| Generic Name | deep brain stimulator for obsessive compulsive disorder (ocd) |
| Applicant |
| Medtronic Inc.,Neurological Division. |
| 7000 central ave., northeast |
| minneapolis, MN 55432 |
|
| HDE Number | H050003 |
| Supplement Number | S278 |
| Date Received | 03/18/2022 |
| Decision Date | 06/15/2022 |
| Product Code | |
| Advisory Committee |
Neurology |
| Supplement Type | normal 75 day track |
| Supplement Reason | change design/components/specifications/material |
| Expedited Review Granted? | No |
| Combination Product | No |
Approval Order Statement Changes to the medtronic model a610 dbs clinician programmer application version 3. 0. The changes consist of a fix to a software defect subject to field corrective action fa1188 reported to fda on september 23, 2021, as well as fixes to other minor defects in the a610 application. |