• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSIST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSIST) BYPASS Back to Search Results
Device Problems Pumping Problem (3016); Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/20/2020
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac trans plantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Medtronic was made aware of this event through a search of literature publications.It was not possible to ascertain specific device information from the literature publication or to match the event with previously reported events.This information is based entirely on journal literature.This event occurred outside the u.S.All information provided is included in this report.Patient information is limited due to confidentiality concerns.Multiple patients and multiple manufacturers were noted in the article; however, a one to one correlation could not be made with unique device serial numbers.The baseline gender/age of the patients represented in the article is male/46 years old.Without a lot number or device serial number, the manufacturing date cannot be determined.Since no device id was provided, it is unknown if this event has been previously reported.Request for additional information will be made and upon receipt a supplemental report will be submitted accordingly.Referenced article: outcomes of patients after successful left ventricular assist device explantation: a euromacs study.Esc heart failure, march 2020; doi: 10.1002/ehf2.12629 investigation of this event is pending and a supplemental report will be sent upon its completion.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A journal article was reviewed which contained information regarding ventricular assist devices (vads).The article discussed clinical outcomes for patients who experienced myocardial recovery and underwent left vad explantation, based upon data from the european registry for patients with mechanical circulatory support (euromacs) clinical registry.Multiple patients and multiple manufacturers were noted in the article; however, a one-to-one correlation could not be made with unique device serial numbers.It was noted that device malfunctions occurred, which were not characterized.All patients later underwent vad explantation for myocardial recovery.The location of the vads is unknown.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A supplemental report is being submitted for investigation completion.Product event summary: pumps with unknown serial numbers were not returned for evaluation.Log file analysis could not be performed since log files covering the reported event date were not available for analysis.As a result, the reported events could not be confirmed due to insufficient evidence.Based on the investigation conducted, there is no evidence to suggest that a device malfunction caused or contributed to the reported events.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
A supplemental report is being submitted for a correction.Concomitant product was added.The device code was corrected from c62860 to c64343.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key10001516
MDR Text Key192680276
Report Number3007042319-2020-02980
Device Sequence Number1
Product Code DSQ
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,l
Type of Report Initial,Followup,Followup
Report Date 05/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2020
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Date Manufacturer Received05/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN ICD
Patient Outcome(s) Other;
Patient Age46 YR
-
-