Catalog Number 8065752438
Device Problem Defective Device (2588)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/08/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported that the probe wasn¿t cutting effectively during a procedure.The product was replaced and procedure completed with no patient harm.
 
Manufacturer Narrative
One opened probe was received with no tip protector, in a zip top bag, for the report of was not cutting effectively during surgery.At the time of receipt it was observed that the probe needle was completely broken off of the rest of the probe assembly.The returned sample was visually inspected and found to be non-conforming with the probe needle completely broken off at the stiffener sleeve, confirming the received condition of the probe.Functional testing, disassembly activities, and a wear evaluation were unable to be performed due to the visual condition of the probe.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The complaint sample was received with the needle completely broken off of the probe assembly.Therefore, a confirmation of the reported event could not be determined and the exact root cause of the complaint could not be verified.The exact root cause of the broken needle cannot be determined from this evaluation.The most likely root cause is handling at any point after the probe was shipped from the manufacturing site.Confirmation of the reported event could not be determined and an exact root cause could not be verified, therefore, no specific action with regard to this complaint was taken.Probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10005451
MDR Text Key189573791
Report Number1644019-2020-00226
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number8065752438
Device Lot Number2351244H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2020
Date Manufacturer Received07/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-