There was no known patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Through follow-up communication livanova learned that the device was cleaned regularly per the instruction for use and that the device is placed outside the operation theatre during use.The device will be sent to the manufacturer site for deep disinfection in order to solve the reported issue.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|