Other relevant device(s) are: product id: evproplus-29us, serial/lot #: (b)(4), ubd: 26-dec-2021, udi#: (b)(4).Product analysis: the delivery catheter system (dcs) was discarded and the valve remains implanted, therefore no product analysis can be performed.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the second tab did not fully release from the delivery catheter system (dcs) upon deployment, resulting in the valve being pulled out of the annulus.The dcs was withdrawn and a new dcs and valve were used to complete the procedure.No additional adverse patient effects were reported. .
|