Model Number SD393.640
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Reporter is synthes sales consultant.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, the two (2) distractor was broken at the sterile processing department.There was no patient involvement.This report is for one (1) distractor for 11mm rod.This is report 2 of 2 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: a product investigation was conducted.Visual inspection: two distractor for 11mm rods (p/n: sd393.640, lot #: h642864) were returned and received.Upon visual inspection, it was observed that both the devices were received with a broken and bent threaded post.On one of the products, the m7x1 threaded post of component 393.640.1 had approximately 3mm broken from the top.One the other, the post was broken just above the base.For one of the products, the broken section was found stuck inside of one of the 393.501 11mm nut assembly components.Dimensional inspection: dimensional inspection of the thread was unable to be performed due to the deformity of the threaded feature.Documentation/specification review: relevant document numbers/revisions were reviewed no nonconformances or documentation changes related to the complaint condition.The revision released after the device produced changed the material to increase its strength which is a design enhancement; no design issues were noted during the investigation.Investigation conclusion: the complaint condition of "broken" was confirmed, as it matches the reported condition, but upon further analysis, it was determined not to be related to the manufacturing process for the following reason: there were no nonconformances during the production of compliant products.The product passed all inspection criteria and was found to be within specifications at the time it was released to the warehouse.The certificate of compliance from supplier avalign certifies that all product was manufactured in accordance with the product drawing sd393_640 rev c, and no non-conformances related to the complaint condition were present for the complainant lot.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: a device history record (dhr) review was conducted: part number:sd393.640, synthes lot number: h642864, supplier lot number: n/a, release to warehouse date: 31jan2019, expiration date: n/a, manufactured by synthes brandywine.Review of the device history record showed that there are potential issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTRACTOR FOR 11MM ROD
Type of Device
TRACTION,APPARATUS,NON-POWERED
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10012215
MDR Text Key190349470
Report Number2939274-2020-02131
Device Sequence Number1
Product Code HST
UDI-Device Identifier10886982006608
UDI-Public(01)10886982006608
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 04/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/29/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSD393.640
Device Catalogue NumberSD393.640
Device Lot NumberH642864
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2020
Date Manufacturer Received06/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-