Manufacturer's investigation conclusion: the reported event of reduced flow followed by the system stopping was not confirmed.The returned centrimag console (serial number (b)(6)) was tested by the european distribution center and by the ppe department.The console was connected to the returned centrimag motor (serial number (b)(6)), was found to perform as intended.The motor was set to various speeds, and the observed flow value correlated accordingly.The reported event of the motor stopping was unable to be reproduced.Although the console operated as intended, the console was observed to have various damages upon arrival and had a dull, flickering display.Since 1st generation centrimag consoles are no longer able to be fully repaired, the console was scrapped per procedure.The root cause of the reported event was unable to be determined through this analysis.Section 9.3, entitled ¿recommended preventive maintenance¿ within the centrimag primary console operating manual instructs users regularly inspect the condition of the centrimag console, and to return to the console for servicing if any damage is observed.Section 10, entitled "emergency/trouble shooting" of the centrimag primary console operating manual states: "the recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." review of the device history record for centrimag primary console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and quality assurance specifications.No further information was provided.The manufacturer is closing the file on this event.
|