Catalog Number 8065751657 |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 04/21/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
A surgeon reported to a company representative that the trocar cannula's valve became dislodged inside a patient's eye during a procedure.There was no harm to the patient.Additional information was received from the surgeon who indicated one of the valve flaps stuck to an instrument and when she removed the instrument, the flap got stuck inside the trocar.She used forceps to remove the loose flap.This is one of two reports for this surgeon.
|
|
Manufacturer Narrative
|
Additional information provided in h.10.Correction provided in b.2.The initial decision to report was incorrect resulting in the initial report being submitted in error.This report does not meet criteria for reporting as a device malfunction.A valve flap sticking to an instrument, getting stuck inside the trocar followed by removal with forceps does not pose a potential risk or likelihood or injury if it were to recur.The manufacturer internal reference number is: (b)(4).
|
|