Model Number V60
Device Problems Overheating of Device (1437); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6).Date of this report: 07-may-2020.
 
Event Description
The customer reported that the unit was not out of gas and the motor was hot upon booting the unit on.There was no patient involvement.The manufacturer¿s international service technician evaluated the ventilator and confirmed a blower failure.
 
Manufacturer Narrative
G4: 20may2020.B4: 21may2020.The customer declined philips repair, so no service was rendered.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10036526
MDR Text Key190522502
Report Number2031642-2020-01669
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-