Model Number CYF-V2
Device Problem Material Perforation (2205)
Patient Problem No Patient Involvement (2645)
Event Type  Malfunction  
Manufacturer Narrative

The device was returned to the service center for evaluation. During the visual inspection, the bending section cover was found damaged. Functional testing was performed finding no issues. The reported event was confirmed. The most likely cause of the damaged bending section was due to unintended use error of the device.

 
Event Description

Olympus received a complaint from a user facility stating that during reprocessing, the bending section covering was found to be ruptured. There was no patient involvement. Additional information could not be obtained.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameVIDEOSCOPE "CYF-V2", EUROPEAN VERSION
Type of DeviceCYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10049319
MDR Text Key222564329
Report Number8010047-2020-02666
Device Sequence Number1
Product Code FAJ
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type USER FACILITY
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/07/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received05/12/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator OTHER
Device MODEL NumberCYF-V2
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer04/17/2020
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/08/2020
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-