Model Number URF-P6R
Device Problems Break (1069); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/21/2020
Event Type  malfunction  
Manufacturer Narrative
Device has been returned for evaluation.The customer¿s complaint was confirmed.A visual inspection on the received condition noted the bending section skeleton tab was protruding from the bending section cover at the insertion tube side.The bending section cover was removed and the skeleton was noted detached/broken with no signs of any sharp edges.There were excessive broken fibers on the image due to the broken bending section skeleton.The instruction manual states the following; ¿do not twist or bend the bending section with your hands¿.Additionally, the instructions for safe use manual indicate that, damage to the bending section would happen if excessive force was applied while angulating in the opposite direction if there was no movement from the bending section; this would occur more so in a narrow environment.Based on similar reported complaints, the most likely cause of the reported event can be attributed to user handling or technique.
 
Event Description
The customer reported to olympus that during reprocessing, a failed leak test occurred.The intended procedure was completed.There was no patient injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The following sections were updated: d4, g4, g7, h2, h4, h6 and h10.Warnings and cautions in the addendum ¿instructions for safe use¿ include a way of handling the endoscope to prevent damages in the bending tube.If the user follows the instruction, damages in the bending tube can be decreased.A review of the device history record found no deviations that could have caused or contributed to the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10057070
MDR Text Key199137997
Report Number8010047-2020-02702
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2020
Was the Report Sent to FDA? No
Date Manufacturer Received12/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-