Model Number 4415 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Per the manufacturer's technical support specialist, there was a blanket flow light that was coming on.During his trouble shooting he saw the cpg light was on and the three speed lights came on.Per the user facility's biomedical technician, he found the solenoid valve to be shorted causing the f4 fuse to be blown and the k1 microcontroller relay to be defective.
|
|
Event Description
|
It was reported that the unit could not start in cardioplegia (cpg) mode and was alarming.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed via information received from the manufacturer's subsidiary.No parts were returned for further evaluation.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|