|
Catalog Number UNKNOWN |
Device Problems
Structural Problem (2506); Device Tipped Over (2589); Detachment of Device or Device Component (2907); Unintended Movement (3026)
|
Patient Problems
Great Vessel Perforation (2152); Anxiety (2328); Depression (2361)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturers ref# (b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.Summary of investigational findings the reported allegations have been further investigated based on the information provided to date.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Rpn/lot# is unknown but there is no evidence to suggest that this device was not manufactured according to specifications.Cook medical will continue to monitor for similar events.
|
|
Event Description
|
Description of event according to initial reporter: it is alleged that "on or about (b)(6) 2010, [pt] was implanted with a cook celect filter.[pt]'s cook celect filter subsequently malfunctioned and caused injury and damages to [pt].Among other things, the cook celect filter fractured.Additionally, [pt's] filter perforated through her ivc abutting her vertebral body and her iliac vein".Patient outcome: it is alleged that "[pt] is at risk for future progressive perforations by the celect filter which could further injure adjacent organs, blood vessels, and structures, as well as fracturing of the ivc filter and migration of the celect filter or pieces thereof.[pt] faces numerous health risks, including the risk of death.[pt] will require ongoing medical care and monitoring for the rest of her life.It is unknown if the filter can be retrieved by any means other than an open surgical procedure".
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2010 via the right internal jugular due to deep vein thrombosis (dvt) while on anticoagulation.Patient is alleging migration, vena cava perforation, and fracture.The patient further alleges anxiety and depression.(b)(6) 2018, report from ct (computed tomography): "caval perforation: yes.Grade 3 perforation of 6 o'clock leg to abut l4 vertebral body.Grade 3 perforation 3 o'clock strut to right posterior common iliac vein.Grade 2 perforation of 9 o'clock strut 0.35 cm.Tilt: yes.8.46 degrees of sagittal tilt.No substantial coronal tilt.Apex of filter touches if not extending 1 mm through anterior wall of ivc.Migration: yes.Filter apex is 6.37 cm below the right renal vein confluence.Pertinent negatives: no ivc stenosis.Additional findings: missing strut or struts between 3 and 6 o'clock.".
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: a2, a4, b1, b5, b6, b7, h6.Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: tilt, anxiety, depression.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported anxiety and depression are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|