DEPUY ORTHOPAEDICS INC US UNKNOWN HIP FEMORAL STEM Back to Search Results
Catalog Number UNK HIP FEMORAL STEM
Device Problem Loss of Osseointegration (2408)
Patient Problems Bone Fracture(s) (1870); Unspecified Infection (1930); Inadequate Osseointegration (2646); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event Description
The literature article entitled, "cementless bipolar hemiarthroplasty for low-energy intracapsular proximal femoral fracture in elderly east-asian patients: a longitudinal 10-year follow-up study" written by masanori nishi, md, ichiro okano, md, takatoshi sawada, md, natsuki midorikawa, md, and katsunori inagaki, md, phd published by hip & pelvis accepted by publisher 07-august-2019 was reviewed.The article's purpose was to assess long-term outcomes of cementless hemiarthroplasties (ha) in elderly east-asian patients with intracapsular proximal femoral fractured treated with cementless ha.Data was compiled from 135 patients and consisted of patients were alive and not alive (implant related deaths not reported within article) at 10 years post implantation.It is noted that only 3 depuy aml stems were utilized amongst non-depuy stems.The article does not clarify which adverse events are associated with specific product brands, and the article does not provide adequate information to determine accurate quantities.The article references "bipolar head" and therefore it is assumed that the depuy stems were utilized in conjunction with depuy bi-polar heads.Figures 1 and 3 provide radiographic images with patient identifiers but does not identify the implants (figure 1 a (b)(6) year old female with osteolysis 15 years post ha; figure 2 is the figure 1 patient who received revision with tha with narrative description providing details that revision was performed 15 years post ha due to bipolar head migration and painful osteolysis; figure 3 (b)(6) year old female with fracture after a fall 11 years post ha implantation).As the implant identity is not specified, the adverse events in radiographic images are captured within the listed adverse events.Depuy products: aml uncemented stem ("fully cylindrical and coated"), bi-polar femoral head (assumed).Adverse events: septic loosening (treated by revision).Dislocation (treated by closed reduction).Intraoperative fracture vancouver type a:3, c:1 (treated by cerclage wiring).Post operative vancouver type a and b1 (type a treated by conservative methods and type b1 internal fixation with a plate - also noted in figure 3).Figure 1 radiographically detected osteolysis on acetabulum (treated with revision with tha) and narrative description indicates associated pain.Figure 3 post operative vancouver type b1 periprosthetic fracture post fall (treated with orif).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.This complaint was opened to document complaints derived through a journal article review.Follow-ups were done to try and obtain additional information from the author of the journal article.No further information was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL STEM
Type of Device
HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10061668
MDR Text Key191138229
Report Number1818910-2020-12186
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 05/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL STEM
Was Device Available for Evaluation? No
Date Manufacturer Received05/28/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-