Model Number CI-1500-04 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Cyst(s) (1800); Discharge (2225)
|
Event Type
Injury
|
Event Description
|
The recipient reportedly experienced chronic intermittent drainage since 2019, due to a cholesteatoma in the mastoid and ear canal.No infection was noted.The recipient was explanted.The recipient was not reimplanted.The recipient has healed.
|
|
Manufacturer Narrative
|
It was reported that the recipient's op-report listed an infection.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.Additional treatment details will not be provided.The external visual inspection revealed that the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|