Catalog Number SH75-5040
Device Problems Material Separation (1562); Use of Device Problem (1670); Device Handling Problem (3265)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 04/23/2020
Event Type  Injury  
Manufacturer Narrative
The actual device concerned was not returned, the device history records (dhr) of the device concerned was reviewed.The production lot, to which the device concerned belongs, passed all in-process inspections including the shaft-pressurized test and the balloon-wrapping test for every product, and the finished product inspections including the shaft tensile strength test and the repetitive balloon inflation/deflation test on representative samples based on sampling plan.No nonconformity or abnormality in the manufacturing processes of the device concerned was found.Probable cause(s) and our comment: the balloon may have damaged by the hardened sharp edge and should have been ruptured when the balloon was inflated.While removing it from the patient, the bulky portion of the ruptured balloon was caught on the sheath and could not be pulled back any more.Further attempts to forcibly pull back the device resulted in breakage of the balloon and the tip, and the broken pieces remained in the patient's blood vessel.
 
Event Description
Shiranui hp is an otw-type pta balloon catheter compatible with 0.018 inch guidewire.Shiranui hp has no approval in usa.However, we intend to report this case as the event occurred on the similar device for crosstella otw" (otw type pta balloon dilatation catheter,0.018" guidewire compatible) in us under 510(k)# k160004." the balloon rupture occurred when the balloon was repeatedly inflated.When it was pulled out, the tip of the catheter was broken.The tip of the catheter with distal part of the balloon broke have remained inside the body and it was removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHIRANUI HP
Type of Device
SHIRANUI HP
Manufacturer (Section D)
KANEKA CORPORATION
2-3-18
osaka, 530-8 288,
JA  530-8288,
Manufacturer Contact
tamiji
nakanoshima
kita-ku
osaka, 530-8-288
MDR Report Key10064348
MDR Text Key193306133
Report Number3002808904-2020-00009
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 04/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Catalogue NumberSH75-5040
Device Lot NumberSP039287
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/24/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
-
-