(b)(4).Summary of root cause analysis: received 2 picture of 2pcs capillary hub of introcan safety-w pur 24g, 0.7x19mm-ap and 1 empty peel pack.From the picture received 1 capillary hub was used sample while the others one seems like unused sample.From the comparison the used capillary hub have shorter length of the capillary compare to the unused capillary.The cut off part of capillary cannot be seen in the picture.The surface of the torn off areas was not visible on the picture.The in-line test equipment is subject of a frequent calibration and a regular verification related to its proper function.Herewith potential malfunctions of the systems would be detected in-time and would be mitigated.Beside the automated 100% in-line test equipment, independent in-process quality controls and final controls on a random sample basis will be conducted by different teams on a regular basis within the production process.Herewith a systematic product defect would be detected.Cut off capillary most likely not appear to be attributed by manufacturing process as the defect is able to be detected and reject by the in-line vision system.Damages induced after assembly process is not possible since the catheter had been protected with protective cap.Since no defect samples provided, no further investigation can be carried out thus this complaint will be concluded as not confirmed.Reviewed the device history record and no abnormalities found during in process and final control inspection.This report is being filed for an item number that is not sold in the united states, however similar items are sold in the united states by b.Braun medical, inc.
|