Qn#(b)(4).The customer returned a sheath extension assembly with dilator inserted for evaluation.The dilator was inadvertently bent during preliminary inspection; however, this did not impact the outcome of this investigation.No defects or anomalies were detected during visual inspection.The hemostasis valve opening inner diameter measured 0.146", which is within the specifications of 0.144"-0.147" per hemostasis valve cap drawing.The dilator hub distal base measured 0.14925", which is within the specifications of 0.149"-0.151" per dilator drawing.The outer diameter of the sheath ext.Assembly measured 0.10225", which is within specifications of 0.099"-0.103" per sheath ext.Assembly graphic.The inner diameter of the sheath tip measured 0.071", which is within specification of 0.070"-0.072" per sheath ext.Assembly graphic.The total length of the sheath body measured 9.625" which is within specifications of 8.875"-9.75" per sheath ext.Assembly graphic.To functionally test the two components, the returned dilator was inserted into the returned sheath hub.The dilator snapped into place and was secure within the sheath hub.No fault found.The returned dilator was able to advance and retract through the returned sheath with minimal resistance.A device history record review was performed on the sheath and dilator with no relevant manufacturing findings identified.The instructions for use (ifu) provided with this kit instructs the user, "ensure dilator hub fully snaps into sheath cap." the reported complaint that the dilator does not lock into the sheath could not be confirmed through testing of the returned sample.The returned components were functionally tested and the dilator locked into the sheath with no issues.The device passed all functional and dimensional testing.Device history records were reviewed with no evidence to suggest a manufacturing related cause.Based on the sample returned no problem was found on the sample.Teleflex will continue to monitor and trend on complaints of this nature.
|