ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR SCS IPG Back to Search Results
Model Number 3660
Device Problems Wireless Communication Problem (3283); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem Inadequate Pain Relief (2388)
Event Date 04/17/2020
Event Type  Injury  
Manufacturer Narrative

The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.

 
Event Description

It was reported after meeting with the patient, the abbott representative attempted to connect to the patient's scs ipg multiple times using the clinician programmer (cp), but the cp displayed "a problem was found with the generator that cannot be repaired" message. The patient stated she had an elbow procedure about six weeks prior and surgery mode was not turned on prior to the surgery. Troubleshooting and review of the device logs indicated the ipg was in the service application or unresponsive state. Surgical intervention may be necessary to replace the patient's scs ipg.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROCLAIM 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of DeviceSCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key10067981
MDR Text Key191384435
Report Number3006705815-2020-01900
Device Sequence Number1
Product Code LGW
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Remedial Action Other
Type of Report Initial,Followup,Followup
Report Date 07/20/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received05/18/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date08/27/2021
Device MODEL Number3660
Device Catalogue Number3660
Device LOT NumberA000085767
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/16/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured08/28/2019
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/06/02/2017/001-C

Patient TREATMENT DATA
Date Received: 05/18/2020 Patient Sequence Number: 1
-
-