Model Number 5392
Device Problem Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Date 05/19/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the top display on the external pulse generator (epg) had damage.The epg is expected to be returned for repair.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: analysis could not confirm the customer comment that the display of the external pulse generator ( epg) was damaged and cracked, however it was noted that the liquid crystal display (lcd) was bleeding.It was noted that the output connector assembly, upper case and lower case were broken ant the keypad flex was damaged.The display wires were pinched ( not compromised) and two case screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the display was cracked.The product has been returned for service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10072545
MDR Text Key191550249
Report Number2182208-2020-00923
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/05/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-