STERIS MEXICO, S. DE R.L. DE C.V. 20" CENTURY STERILIZER Back to Search Results
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2020
Event Type  malfunction  
Manufacturer Narrative
A steris service technician arrived onsite to inspect the sterilizer and found that the unit had alarmed "water in chamber" and observed water in the chamber.During his inspection, the technician found that the ck3 check valve required replacement.As the ck3 check valve remained open, this allowed excess water to enter and leak from the chamber resulting in the reported event.The century sterilizer operator manual states the following instructions if a "water in chamber" alarm occurs (8-12), "1.Do not open door 2.Silence alarm.3.Call service immediately." the technician replaced the ck3 check valve, tested the unit, confirmed it to be operating according to specifications, and returned it to service.No additional issues have been reported.
 
Event Description
The user facility reported that water was leaking onto the floor from their 20" century sterilizer.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
20" CENTURY STERILIZER
Type of Device
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, oh 
3927453
MDR Report Key10075507
MDR Text Key199315427
Report Number3005899764-2020-00033
Device Sequence Number1
Product Code FLE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/20/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-