• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-290SL
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/27/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device in this report has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed from the facility that in unspecified timing the scope communication error b30 which indicated that there was the communication problem between the endoscope and video processor occurred at the user facility.Other detailed information was not provided.There was no report of patient injury associated with the event.
 
Manufacturer Narrative
This supplemental report is being submitted to additional information.Olympus medical systems corp.(omsc) reviewed the manufacture history (dhr) of the device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However there was the possibility that the reported phenomenon was attributed to the contact failure of the output socket due to wear, and/or the other devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10077746
MDR Text Key209001912
Report Number8010047-2020-02826
Device Sequence Number1
Product Code GCT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberCLV-290SL
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-