Model Number 1012453-20
Device Problems Deflation Problem (1149); Difficult to Remove (1528); Material Separation (1562); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/06/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4). The device is expected to be returned for evaluation. It has not yet been received. A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was performed to treat a lesion in the moderately calcified left anterior descending coronary artery. A 4x20mm nc trek balloon dilatation catheter (bdc) was advanced without resistance, and the balloon was inflated once to 24 atmospheres. However, the balloon only partially deflated within the unspecified stent. The partially deflated balloon was retracted into the guide catheter with resistance and force applied. The shaft of the bdc separated, so the whole system was removed. A non-abbott balloon was used to successfully complete the procedure. There were no adverse patient effects and no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC TREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10079390
MDR Text Key193363080
Report Number2024168-2020-04457
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648152085
UDI-Public08717648152085
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K103153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/26/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1012453-20
Device Catalogue Number1012453-20
Device Lot Number00121G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-