• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751767
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 05/08/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A surgeon reported after finishing preparation work, it was observed that the cassette leak when tested.The product was replaced and the planned procedure was completed.There was no patient involvement and no harm to the operating room staff.
 
Manufacturer Narrative
Additional information provided in h.6.And h.10.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The customer reported cassette leakage.A sample was not received to date for this complaint report; therefore, visual inspection or functional testing could not be conducted.Without a sample it cannot be determined if there were any physical anomalies that led to the customer's experience.The root cause of the customer's complaint could not be established as a sample has not been received.Without a sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship, if any, of the device to the reported incident cannot be determined.No action will be taken for this occurrence, as the root cause is unknown and a sample was not returned.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Additional information provided in d.10., h.3., h.6., and h.10.The returned sample was visually inspected and no obvious defects were found.The non-invasive flow sensor (nifs) on the cassette housing was in good condition.A calibrated console representing the current software version was used to test the sample.The ball in the drip chamber¿s check valve moved freely per specification.The light emitting diode (led) rings on the console turned green as the probe connectors were engaged to the console indicating the proper communication between the probe and the console.The submerge leak test was performed on the cassette.No leakage was observed.The sample could prime and pass intraocular pressure (iop) calibration successfully.No anomalies were observed during priming.No system message code was generated during testing.Fluid flowed from the balanced salt solution (bss) bottle to the drain bag without any interfering.No leakage was detected from the pump elastomer or on the pump area of the fluidics module.Cleaning process was able to be performed after functional test had completed.The root cause of the customer's complaint could not be established; the returned sample functioned per specifications.After an investigation of this complaint, it has been determined that this sample functioned per specifications; therefore, no corrective action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key10080967
MDR Text Key192024704
Report Number1644019-2020-00254
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Catalogue Number8065751767
Device Lot Number2283482H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2020
Date Manufacturer Received11/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM
-
-