• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Date 04/29/2020
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) was not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.The getinge service territory manager (stm) who encountered the issue replaced the backplane board as it had no audio alarm.The stm completed the pm and performed all functional and safety tests were passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.(b)(6).
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) that the system failure alarm on the cardiosave intra-aortic balloon pump (iabp) did not have any audio.There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd
mahwah, NJ 07430
MDR Report Key10083469
MDR Text Key192056405
Report Number2249723-2020-00789
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 05/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received04/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-