RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Charging Problem (2892)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of this report: 22 may 2020.
|
|
Event Description
|
The customer reported that the battery cannot be charged.There was no patient involvement.Technical support provided remote assistance to the customer.The customer reported to have replaced the power supply to address the problem, however, the power supply appears to be faulty as there is no direct current (dc) output.Technical support advised the customer to replaced the power supply once again.
|
|
Manufacturer Narrative
|
G4: 03jun2020.B4: 04jun2020.The customer reported to have replaced the power supply and alternate current (ac) inlet filter to address the problem.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|