DEPUY SYNTHES PRODUCTS LLC BATTERY OSCILLATOR II FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Model Number 530.710 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi: (b)(4).
|
|
Event Description
|
It was reported that the battery oscillator device was leaking fluid.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in the surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During evaluation by quality engineering it was determined that the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during the evaluation, it was determined that the device oscillator head was cracked at the head base, internal components were corroded, unknown liquid was found inside of the slide sleeve, and the motor had excessive vibration and drew too much current.It was further determined that the device failed pre-test for general condition, check the saw blade coupling and had a cracked oscillating head base.The assignable root cause was improper maintenance and a design issue for the cracked oscillating saw head which had been evaluated under capa.A device history review was performed which indicated that there were no relevant issues identified during the manufacture of the device that may have contributed to the reported condition.
|
|
Search Alerts/Recalls
|
|
|