The customer reported a high blower temperature alarm.
The ventilator was being used on a patient at the time of the reported event, however, there was no patient harm.
The patient's information was not provided.
The patient had to be moved to a different ventilator as a result of this event.
There was no additional medical intervention being required.
The field service engineer (fse) evaluated the ventilator and confirmed the occurrence of the reported alarm.
The fse installed new air inlet filter, cleaned the cooling fan filter, and verified that the cooling fan motor was operating within specifications.
The ventilator successfully passed functionality testing after repair was completed.
|