There was no patient involvement.
Livanova (b)(4) manufactures the s5 gas blender system.
The incident occurred in (b)(6).
A review of the dhr could not identify any deviations or nonconformities relevant to the issue a field service technician was dispatched to the facility to investigate the device and was not able to reproduce the reported issue.
He tested the unit and when the flow rate of o2+ air fell below the set value the alarm properly worked.
As results of the service activity no abnormalities were found.
Based on the information available, it cannot be ruled out that a user error during the set-up phase led to the reported issue.
|