Qn# (b)(4).The actual device was not returned; however, the customer provided two videos for evaluation.Visual examination of the videos confirmed a leak in the juncture hub during use.However, full visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed and no relevant findings were identified.The instructions-for-use provided with this kit warns the user, "do not apply excessive force in placing or removing catheter.Failure to do so can result in catheter breakage.If placement or withdrawal cannot be easily accomplished, an x-ray should be obtained and further consultation requested." the customer report of a catheter leak was confirmed by visual examination of the customer supplied videos.The videos showed the catheter leaking from an apparent hole in the juncture hub during use.A full complaint verification testing could not be performed as the sample was not returned for analysis.A device history record review was performed and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|