OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA II VIDEO SYSTEM CENTER Back to Search Results
Model Number CV-180
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device has been returned for evaluation.The customer¿s complaint of front panel switches not functioning was confirmed.The service technician observed front panel was damaged.The service technician ensured that the unit was fully functional after the repair.The cause was traced to component failure.No further information was reported.
 
Event Description
The customer reported to olympus during an endoscopic retrograde cholangiopancreatography procedure, the front panel lights remained on and the buttons were not functioning.The image was displayed much smaller, as if it was in picture in picture module instead of full screen.The procedure was completed with the same device.There was no patient injury reported.
 
Manufacturer Narrative
This supplemental report is submitted to provide the results of the legal manufacturer¿s investigation and device history record (dhr) review.Updates to sections h4 and h6.The dhr was reviewed for the subject device.No anomalies were noted and it was verified the device was manufactured in accordance with documented specifications and procedures.The legal manufacturer performed an investigation and was unable to determine a conclusive root cause for the component failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II VIDEO SYSTEM CENTER
Type of Device
VIDEO SYSTEM CENTER
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10106140
MDR Text Key224676794
Report Number8010047-2020-03028
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCV-180
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/31/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-