• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR; SCS IPG Back to Search Results
Model Number 3660
Device Problem No Apparent Adverse Event (3189)
Patient Problem Inadequate Pain Relief (2388)
Event Date 05/20/2020
Event Type  Injury  
Manufacturer Narrative
An elective replacement indicator message was reported to abbott.The device was not returned for analysis; however, logs and/or assessment report were assessed and indicate that the battery voltage level of the device is within specifications to trigger the eri indicator.Based on the information received, the cause of the reported incident is consistent with the battery nearing end of life.
 
Event Description
It was reported that the patient's ipg had reached end of life.In turn, surgical intervention was undertaken wherein the ipg was explanted and replaced, resolving the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM¿ 5 ELITE IMPLANTABLE PULSE GENERATOR
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key10108983
MDR Text Key193313723
Report Number1627487-2020-21232
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020192
UDI-Public05415067020192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/22/2019
Device Model Number3660
Device Catalogue Number3660
Device Lot Number5911006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight73
-
-