Model Number 174025
Device Problem Failure to Fire (2610)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/18/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, while closing the mesenterium after a laparoscopic gastric bypass procedure, the surgeon was unable to squeeze the handle.The stapler got blocked and no staples were released.Another stapler was used to complete the case.There was no patient injury.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: g4, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the e-piece was disengaged on one side and the single use loading unit was received partially applied and preloaded on the instrument.Functional testing found the staple did not deploy properly and turned in the e piece.It was reported that the device did not fire.The reported issue was confirmed.The most likely root cause was traced to manufacturing of the device.A misaligned lower cover allowed the e-piece to be welded in the incorrect position, the lack of clearance between the e-piece and the descending staples could generate a turned clip condition during its firing cycle.After the turned clip condition is seen, if the handle of the unit is squeezed it can cause the e-piece to be partially disengaged; if the end user tries to remove the dlu/cartridge from the unit while having the turned clip condition it can result in a e-piece disengage.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE ENDO HERNIA
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key10111142
MDR Text Key193424386
Report Number2647580-2020-01721
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521070592
UDI-Public10884521070592
Combination Product (y/n)N
PMA/PMN Number
K912097
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number174025
Device Catalogue Number174025
Device Lot NumberP9K1395Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2020
Date Manufacturer Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberR202003-06
Patient Sequence Number1
-
-